Intraobserver and interobserver variance in the measurement of ultrasound-estimated bladder weight
Abstract:
Recently, we reported that ultrasound-estimated bladder weight (UEBW) could be used as a noninvasive tool in evaluating the degree of bladder hypertrophy. This study was designed to test the reproducibility of its measurement. The overall mean UEBW of the initial measurement and that of the second one by one observer in 36 cases was 42.8 +/- 22.6 g and 42.9 +/- 22.6 g, respectively. The paired differences between two measurements ranged from -3.4 to 2.7 g, with a mean difference of only -0.1 g. The interobserver variance was also slight. The paired differences between UEBW measurements derived from the two observers in 32 cases ranged from -6.0 to 7.9 g, with a mean difference of only 0.1 g. The Cochran's test statistic of the measurements of UEBW was 0.142 for one observer and 0.130 for two, which were less than its 5% critical value. Due to these favorable characteristics, UEBW could be reliable tool for the investigation of bladder hypertrophy with a sufficient reproducibility.