MR-guided biopsy using a T2-weighted single-shot zoom imaging sequence (Local Look technique)
A Buecker1, G Adam, J M Neuerburg
1Clinic for Diagnostic Radiology, University of Technology Aachen, Germany. buecker@LINUS.RAD.RWTH-AACHEN.DE
Journal of Magnetic Resonance Imaging : JMRI
|August 14, 1998
Summary
The Zoom Imaging technique, also known as Local Look (LoLo), offers effective T2-weighted imaging for MR-guided percutaneous interventions. It provides excellent lesion contrast and minimal needle artifacts, ensuring accurate needle tip visualization during procedures.
Area of Science:
- Radiology
- Medical Imaging
- Interventional Radiology
Background:
- Magnetic Resonance (MR)-guided percutaneous interventions require precise visualization of instruments and target lesions.
- Traditional imaging techniques may present limitations in terms of speed, contrast, or artifact generation during these procedures.
Purpose of the Study:
- To evaluate the clinical utility of the T2-weighted single shot turbo spin-echo technique, "Local Look" (LoLo), now "Zoom Imaging," for MR-guided percutaneous interventions.
- To compare the performance of the LoLo technique against a fast gradient-echo sequence for needle guidance.
Main Methods:
- Twenty-eight MR-guided percutaneous procedures were performed on 22 patients using a 1.5-T MR system.
- The LoLo technique, acquiring T2-weighted images in 600 ms with a small field of view and single-shot method, was used for all procedures.
- Orthogonal pulse orientation prevented backfolding artifacts; half of the procedures also used a fast gradient-echo sequence for comparison.
Main Results:
- All 28 procedures were technically successful with no complications.
- The LoLo technique demonstrated superior or equal delineation of the needle tip compared to the gradient-echo sequence in 100% of cases.
- Images acquired with LoLo exhibited good lesion contrast and smaller needle artifacts than the gradient-echo technique.
Conclusions:
- The LoLo (Zoom Imaging) technique is an accurate and effective method for MR guidance of percutaneous procedures.
- It provides excellent lesion contrast and minimal needle artifacts, crucial for precise needle tip monitoring.
- Gradient-echo sequences may be beneficial for planning complex cases, but LoLo is superior for real-time needle tip visualization.