Comparing different lots of activated partial thromboplastin time reagent: analysis of two methods
1Department of Medical Education, Abbott Northwestern Hospital, Minneapolis, Minnesota 55407, USA.
American Journal of Clinical Pathology
|August 15, 1998
Summary
Comparing activated partial thromboplastin time (APTT) reagent lots revealed significant variability. Lot-to-lot differences impact unfractionated heparin (UH) monitoring, suggesting APTT may not be optimal for therapy management.
Area of Science:
- Clinical Chemistry
- Hematology
- Laboratory Medicine
Background:
- Activated partial thromboplastin time (APTT) is crucial for monitoring unfractionated heparin (UH) therapy.
- Variability in APTT reagent performance can affect therapeutic decisions.
- Assessing lot-to-lot consistency is essential for reliable laboratory diagnostics.
Purpose of the Study:
- To compare the performance of two different lots of APTT reagent from the same manufacturer.
- To evaluate the impact of reagent lot variability on APTT results and therapeutic ranges for UH.
- To assess the agreement between different APTT reagent lots and within-lot duplicates.
Main Methods:
- Analysis of 200 plasma samples using two APTT reagent lots on the same instrument.
- Comparison with 267 additional plasma samples analyzed in duplicate with a single reagent lot.
- Application of regression analysis and Bland-Altman assessment of agreement on logarithmically transformed data.
Main Results:
- Suboptimal agreement was observed with regression analysis, particularly at higher APTT values.
- The new reagent lot yielded results that were 88.9% of the old lot's values, narrowing the UH therapeutic range.
- Duplicate APTT determinations were 97.6% of the initial value, with 95% limits of agreement of +/-10% within the therapeutic range.
Conclusions:
- Significant variability exists between APTT reagent lots, impacting the therapeutic range for unfractionated heparin.
- The Bland-Altman method is superior to regression analysis for assessing agreement between APTT reagent lots.
- APTT variability, both within and between lots, may limit its optimization for monitoring UH therapy.