A multicenter evaluation of a new continuous cardiac output pulmonary artery catheter system
F G Mihm1, A Gettinger, C W Hanson
1Department of Anesthesia, Stanford University, CA, USA.
Objective:
To validate a new system of continuous cardiac output monitoring.
Design:
Multicenter, prospective, nonrandomized clinical study.
Setting:
Four university hospitals.
Patients:
Forty-seven adult intensive care unit patients.
Interventions:
Pulmonary artery catheterization.
Measurements And Main Results:
Continuous and bolus cardiac output measurements were obtained over 72 hrs. The 327 continuous cardiac output measurements compared favorably with bolus cardiac output measurements (bias = 0.12 L/min, precision = +/-0.84). The continuous cardiac measurement was not adversely affected by temperatures of <37 degrees C or >38 degrees C, high (>7.5 L/min) or low (<4.5 L/min) cardiac output values, or duration (72 hrs) of the study.
Conclusions:
This continuous cardiac output system provides a reliable estimate of cardiac output for clinical use if applied in conditions similar to this study. The combination of a continuous measure of cardiac output with other continuous physiologic monitoring (arterial and mixed venous oxygen saturation, oxygen consumption, etc.) may provide important information that no single parameter could achieve.
More Related Videos
Related Concept Videos
Cardiac Catheterization II: Right Heart Catheterization
Cardiac Catheterization III: Left Heart Catheterization
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Cardiac Catheterization IV: Nursing Management
Cardiac Catheterization I: Pre-Procedure Overview