Measurement of blood volume using indocyanine green measured with pulse-spectrophotometry: its reproducibility and
1Department of Anesthesiology, Osaka University Medical School, Suita City, Japan. YLHE@anes.med.osaka-u.ac.jp
Objective:
To systematically investigate the reproducibility and reliability of a newly developed, less invasive approach of estimating blood volume (BV), using indocyanine green (ICG) measured with pulse-spectrophotometry.
Design:
Prospective, clinical study.
Setting:
Surgical unit at a university hospital.
Patients:
Twenty-two patients undergoing general anesthesia for elective surgery and seven healthy volunteers.
Interventions:
Catheters were inserted into the forearm veins of healthy volunteers for the administration of ICG and blood sampling for the measurement of hemoglobin concentration.
Measurements And Main Results:
The distribution volumes of ICG in seven healthy volunteers were estimated repetitively following three or four consecutive intravenous administrations at 30-min intervals. A low intrasubject coefficient of variation of 3.94 +/- 2.03 (SEM) % and a reasonable intersubject coefficient of variation of 13.3 +/- 5.52% (in mL/kg) for the BV measurements were obtained. In addition, ICG was administered to 22 patients, first under general anesthesia by a bolus, and then by a bolus with a constant-rate infusion. The ICG blood concentration was noninvasively measured with pulse-spectrophotometry. The blood concentration time courses following both bolus and constant-rate infusion were well fitted by the one-compartment model, indicating that the distribution equilibrium of ICG is instantaneous. The distribution volumes estimated following bolus injection correlate closely with the distribution volume estimated based on constant-rate infusion administration (r2 = .90).
Conclusions:
The BV estimation with a bolus injection of ICG and pulse-spectrophotometry is reliable, as reflected by the reproducible BVs estimated in the same subject. The integrated pulse-spectrophotometry monitoring system offers a less invasive and useful tool for bedside estimation of BV.
More Related Videos
05:14Characterization of Vascular Morphology of Neovascular Age-Related Macular Degeneration by Indocyanine Green Angiography
Published on: August 11, 2023
09:18Doppler Ultrasound-Based Leg Blood Flow Assessment During Single-Leg Knee-Extensor Exercise in an Uncontrolled Setting
Published on: December 15, 2023
Related Concept Videos
Measurement of Bioavailability: Pharmacokinetic Methods
Measurement of Bioavailability: Pharmacodynamic Methods