Differences in observer variability of ultrasound measurements of the proximal and distal abdominal aorta
H J Pleumeekers1, A W Hoes, P G Mulder
1Erasmus University Rotterdam, The Netherlands.
Objective:
To assess the observer variability of ultrasound measurements of the abdominal aorta and to study whether observer variability is influenced by the site of measurement or cardiovascular risk factors.
Setting:
Population based screening programme for abdominal aortic aneurysms.
Methods:
For 135 subjects taking part in a screening programme for abdominal aortic aneurysms, two of the three ultrasonographers measured the distal and proximal ultrasound diameter of the abdominal aorta, using B-mode ultrasound, according to the Rotterdam study scanning protocol.
Results:
The mean difference between two different observers was 0.06 mm (95% CI-0.15 to 0.27) for measurements of the distal aorta and 0.32 mm (95% CI 0.09 to 0.55) for the proximal aorta. Maximal differences between observers for measurements of both the distal and proximal aortic diameter were 4.0 mm. Interobserver variability in the proximal and distal measurements of the abdominal aorta was not related to the level of the major cardiovascular risk indicators. However, interobserver variability in ultrasound measurements of the proximal aorta increased with increasing waist circumference and increasing diameter of the proximal aorta.
Conclusion:
Ultrasonographic readings of the distal and proximal aortic measurements can be interpreted within a range of plus or minus 3 mm. Ultrasound measurements are more accurate for the distal than for the proximal measurement. Definition of the aortic diameter based on a combination of both distal and proximal measurement may be more accurate.
Related Concept Videos
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessment of blood pressure in brachial artery(two-step method)
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.