H Deriaz1, E Baras, M Bunodière
1Département d'anesthésie-réanimation, hôpital Saint-Antoine, Paris, France.
Annales Francaises D'Anesthesie Et De Reanimation
|January 1, 1997
Summary
Monochromatic infrared gas spectrometers (MIS) can display incorrect volatile agent (VA) concentrations due to erroneous selection. This study highlights dangerous underestimations and overestimations, emphasizing the need for careful device verification during anesthesia.
Area of Science:
- Anesthesiology
- Medical Device Technology
- Patient Safety
Background:
- Monochromatic infrared gas spectrometers (MIS) calculate volatile agent (VA) concentration using IR absorption and a specific gain factor.
- These devices cannot identify the actual VA present, leading to potential discrepancies between displayed and actual concentrations.
- While desflurane vaporizers prevent misfilling, erroneous VA selection on the MIS remains a risk.
Purpose of the Study:
- To evaluate the displayed concentrations on a Capnomac Ultima (Datex) monitor following incorrect volatile agent (VA) selection.
- To categorize the potential errors arising from misidentifying anesthetic agents.
Main Methods:
- Desflurane, isoflurane, or enflurane were delivered at constant concentrations.
- The MIS successively measured and displayed concentrations for desflurane, sevoflurane, isoflurane, enflurane, and halothane.
- Displayed concentrations were compared against actual vapor concentrations.
Main Results:
- Errors were categorized into dangerous (displayed concentration much lower than actual, e.g., desflurane/sevoflurane misselection), evident (displayed concentration >10 vol%), and uncomfortable (displayed and actual concentrations similar, e.g., isoflurane instead of desflurane).
- Dangerous errors occurred when desflurane or sevoflurane were erroneously selected.
- Uncomfortable situations arose when isoflurane was selected instead of desflurane, requiring careful device checking.
Conclusions:
- Erroneous volatile agent (VA) selection on monochromatic infrared gas spectrometers (MIS) can lead to significant and potentially dangerous miscalculations of anesthetic agent concentrations.
- The risk of misidentification necessitates vigilant monitoring and verification of device settings by clinicians.
- Implementing strict checking protocols for MIS settings is crucial for patient safety in anesthesia delivery.