Validation of an enzyme calibrator--an IFCC guideline. International Federation of Clinical Chemistry
Related Experiment Video
Updated: Oct 28, 2026

Detection of Antibodies That Neutralize the Cellular Uptake of Enzyme Replacement Therapies with a Cell-based Assay
Published on: September 10, 2018
1Laboratoire de Biochimie Appliquée, Faculté de Pharmacie, Université Louis Pasteur, Illkirch, France.
Objectives:
The objective of this guideline is to improve standardization in clinical enzymology in order to improve intermethod comparability of patients' results.
Design And Methods:
The reference system, combination of the reference method and the reference material, is used to produce a reference value for a given catalytic activity. Sets of methods are formed of methods exhibiting the same analytical specificity. Materials intended to be used as enzyme calibrators are experimentally checked for their commutability.
Results:
The transfer of accuracy from the reference value to patients' results is dependent on methods (analytical specificity) and on materials (experimentally assessed commutability). The feasibility of this approach was demonstrated with materials of high level for several enzymes and for each of them for several routine methods.
Conclusion:
Expected advantages of this approach in clinical enzymology are presented.
Related Concept Videos
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Data Validation
Key parameters for method validation include:
Glassware Calibration
Volumetric flasks: Volumetric flasks are designed to prepare aqueous solutions of precise volumes accurately with a calibration line on the neck. To calibrate a volumetric flask, it is important to fill it with distilled...
EDTA: Indirect and Alkalimetric Titration