Related Experiment Video
Updated: Aug 14, 2026

Expired CO2 Measurement in Intubated or Spontaneously Breathing Patients from the Emergency Department
Published on: January 29, 2011
Can capnography detect bronchial flap-valve expiratory obstruction?
1Department of Anesthesiology, University of California-Irvine, UCI Medical Center, Orange 92868, USA.
Objective:
We have previously shown in a mechanical lung model [1] that bronchial flap-valve expiratory obstruction results in sequential lung expiration, best detected by prolonged and low magnitude tracheal expired flow (V) from the obstructed lung. However, the normal expiratory resistance of clinical ventilation circuits might also generate prolonged, low value exhaled V, that could be confused with bronchial flap-valve obstruction. We reasoned that bronchial flap-valve obstruction would also cause sequential CO2 unloading from each lung and result in a biphasic tracheal capnogram.
Methods:
To test this hypothesis, we ventilated (VT, 650 ml; f, 10 br/min) a dual mechanical test lung, with each side connected to a separate alcohol-burning chamber. An airway adapter-monitor system measured airway V, P, PCO2, and FO2. The circumference of the diaphragm in a respiratory one-way valve was trimmed to generate unidirectional resistance to expiratory V. Measurement sequences were repeated after this flap-valve was interposed in the left "main-stem bronchus."
Results And Discussion:
During moderate or severe left bronchial flap-valve obstruction, left bronchial V was delayed so that the left lung anatomical dead space (devoid of CO2) mixed with normal right exhalate to depress the expiratory upstroke or early plateau of the tracheal capnogram. During severe obstruction, decreased perfusion of the left lung caused lower alveolar PCO2. Then, prolonged low V from the left bronchus also resulted in depression of the end of the tracheal alveolar plateau. In general, the low magnitude of bronchial V from the obstructed lung limited its effect on the tracheal capnogram and the best marker of sequential lung emptying during bronchial flap-valve obstruction may be late exhaled V without reduction in total tidal volume.
More Related Videos
Related Concept Videos
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...
Respiratory System Abnormal Finding I: Inspection and Percussion
Inspection Findings
During an inspection, several findings may suggest the presence of respiratory distress or disease. Pursed-lip breathing, where exhalation is slowed by...
Respiratory System Abnormal Finding II: Palpation and Auscultation
Palpation Findings
During a respiratory assessment, palpation can reveal several vital abnormalities:
Acute Respiratory Failure-III
Atelectasis II: Pathophysiology