Metallic fragment deposits during phacoemulsification
Related Experiment Video
Updated: Oct 24, 2026

Multi-Scale Modification of Metallic Implants With Pore Gradients, Polyelectrolytes and Their Indirect Monitoring In vivo
Published on: July 1, 2013
J J Martínez-Toldos1, J C Elvira, J R Hueso
1San Juan University Hospital, Alicante, Spain.
Purpose:
To describe the nature and origin of foreign metallic intraocular bodies appearing after phacoemulsification.
Setting:
Department of Ophthalmology, University Hospital of San Juan, Alicante, Spain.
Methods:
Two metallic particles, 1 extracted during surgery and the other from a patient in whom surgery had been performed, were studied by scanning electronic microscopy and X-ray dispersive energy spectroscopy. Also evaluated were all metallic elements used in phacoemulsification including phaco tips, Sinskey hooks, and handpieces. A morphologic analysis at various magnifications was performed and the composition of the elements studied.
Results:
Scanning electronic microscopy showed irregularities of the phaco tip and Sinskey hook after their use. Spectrographic analysis found that the phaco tip was mainly aluminum and titanium; the Sinskey hook, iron, chromium, cobalt, and nickel; the handpiece, iron, chromium, and nickel; and the 2 metallic particles, iron, chromium, and nickel, the same as the handpiece.
Conclusion:
The metallic particles extracted corresponded to those of the phaco handpiece. Vibration during prolonged use of the phacoemulsifier probably caused these particles to chip off the handpiece.
More Related Videos
12:19Identification of Metal Oxide Nanoparticles in Histological Samples by Enhanced Darkfield Microscopy and Hyperspectral Mapping
Published on: December 8, 2015
11:49A Continuous-flow Photocatalytic Reactor for the Precisely Controlled Deposition of Metallic Nanoparticles
Published on: April 10, 2019
Related Concept Videos
Qualitative Analysis
For instance, group IV...
Colloidal precipitates
Coagulation
Extraction: Advanced Methods