Imaging In-Stent Restenosis: An Inexpensive, Reliable, and Rapid Preclinical Model
Published on: September 14, 2009
Coronary stents: In vitro aspects of an angiographic and ultrasound quantification with in vivo correlation
E V Pomerantsev1, Y Kobayashi, P J Fitzgerald
1Stanford-UCSF Health Services, Stanford, Calif, USA.
Background:
The validity of quantitative coronary angiography (QCA) after stent placement has been questioned because the optical density of a metallic stent, added to the density of a contrast-filled lumen, could affect border definition.
Methods And Results:
We deployed 3.0- and 4.0-mm Palmaz-Schatz, Wiktor, Multilink, NIR, and InStent stents in precision-cast phantoms. Central lumens of 2.0 mm were created. There was no difference between the "true" diameters of any stented lumen by both QCA and quantitative ultrasonic (QCU) measurement poststenting. QCA systematic error (SE) varied from 0.01 for the Wiktor stents to 0.14 mm for the Palmaz-Schatz stents; the random error (RE) was 0.03 to 0.14 mm. QCU SE varied from 0.05 to 0.11 mm, and RE ranged from 0.01 to 0.07 mm. At the next stage, 4.0-mm Wiktor and Palmaz-Schatz stents were deployed into the phantom lumens; 1.5-, 2.0-, 2.5- and 3.0-mm lumens were created inside the stents. QCA and QCU measurements of 1.5- to 2.5-mm residual lumens were overestimated by 0.1 to 0.3 mm. In the 3. 0-mm residual lumen within the Wiktor stent, QCA underestimated the luminal size by -0.1 mm. There was no QCA inaccuracy for a 3.0-mm lumen within the Palmaz-Schatz stent. In patients, in 25 stented segments in both the Palmaz-Schatz and Wiktor groups, there was no difference between QCA and QCU diameters.
Conclusions:
QCU is sufficiently precise for the assessment of the coronary lumen after stenting. QCA can be used as an accurate method of poststent assessment, except when a very mild recurrence within a highly opaque stent is measured. In that instance, QCA may underestimate the luminal diameter.
More Related Videos
05:04Implantation of Human-Sized Coronary Stents into Rat Abdominal Aorta Using a Trans-Femoral Access
Published on: November 19, 2020
06:18Intravascular Ultrasound Image-Based Finite Element Modeling Approach for Quantifying In Vivo Mechanical Properties of Human Coronary Artery
Published on: December 6, 2024
Related Concept Videos
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion, evaluates...
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for diagnosing...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Imaging Studies VII: Vascular Imaging