Does the hemoglobin concentration in fetal blood interfere with the accuracy of fetal reflection pulse oximetry?
G M Arikan1, M C Haeusler, J Haas
1Department of Obstetrics and Gynecology, University of Graz, Austria. arikan@email.kfunigraz.ac.at
Objective:
Our goal was to investigate the influence of the fetal hemoglobin (Hb) concentration on the accuracy of reflection pulse oximetry.
Methods:
179 fetuses were monitored by a reflection pulse oximetry system (RPOX MK2). Pulse oximetry measurements (SO2POX) were compared with the oxygen saturation and pH of umbilical artery and vein (UA-, UV-SaO2) immediately after delivery. Correlation and linear regression analyses were performed. Because it is unknown whether low or high Hb concentrations might interfere with the SO2POX measurements, the influence of the Hb level on the correlations was evaluated by weighting the linear regression in both ways; by Hb concentrations as weighting factor and as an inverse weighting factor (1/Hb).
Results:
There was a significant correlation between SO2POX and UV-SaO2 (for the last 10 min of delivery, r = 0.45, p < 0.0005) but not with pH or UA-SaO2. Linear regression analysis between SO2POX and UV-SaO2 showed a multiple R of 0.45. Hb concentration in fetal blood ranged from 9.2 to 20.5 mg/dl. The weighted regression indicated a multiple RHb of 0.49, which did not differ by the inverse weighting (multiple Rinverse Hb: 0.49).
Conclusions:
Pulse oximetry measurements seem to reflect oxygen saturation in fetal blood, however further improvement is necessary. The Hb concentration in fetal blood does not interfere with the accuracy of fetal pulse oximetry and need not be taken into consideration in calibration curves.
Related Concept Videos
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Factors Affecting Erythropoiesis
Several factors influence the erythrocyte production rate, with tissue oxygen level being among the most critical. Intense exercise or high altitudes can cause tissue hypoxia, which triggers the kidneys to release more erythropoietin (EPO) into the bloodstream.
EPO then...
Oxygen Transport in the Blood
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...