Modified Headholder and operating table for intra-operative MRI in neurosurgery
Related Experiment Video
Updated: Oct 28, 2026

Magnetic Resonance-Guided Stereotaxy for Infusions to the Pig Brain
Published on: March 31, 2023
C R Wirtz1, V M Tronnier, F K Albert
1Department of Neurosurgery, Ruprecht-Karls-Universität Heidelberg, Germany.
Abstract:
In order to facilitate intra-operative use of magnetic resonance imaging (MRI) in neurosurgery an MRI-compatible headholder was developed and adapted to a modified MR-couch simultaneously serving as tabletop for the operating table. To allow shock-free transport into the scanner the wheels of the operating table were replaced by an air cushion mechanism. In 75 procedures the system proved to be reliable and safe. Image quality was not impaired by the fixation device. With growing routine the transfer became straightforward, requiring approximately 10 min. Intra-operative MRI is thus made possible with minimal changes to the standard surgical environment. Its benefit however, still remains to be critically investigated.