Optimization of shocking lead configuration for transvenous atrial defibrillation
1Electrophysiology Laboratory, Sinai Samaritan, Medical Center, Milwaukee, Wisconsin, USA.
Introduction:
High atrial defibrillation energy requirements (ADER) in patients with chronic atrial fibrillation (AF) may limit the acceptance of transvenous atrial defibrillation. We evaluated an optimized defibrillation electrode configuration that could help to reduce the ADER in patients with AF.
Methods And Results:
We tested ten different configurations in nine dogs with AF (3.33+/-2.92 days) induced by rapid atrial pacing. The configurations were: right atrial (RA) appendage as anode and coronary sinus (CS) as cathode; RA and innominate vein (I) as anode to CS (cathode); RA-CS (anode) to I (cathode); I-CS (anode) to RA (cathode); RA and left lateral subcutaneous patch (P) as anode to CS (cathode); RA-CS (anode) to P (cathode); P-CS (anode) to RA (cathode); superior vena cava (SVC) and CS (anode) to RA (cathode); RA-CS (anode) to SVC (cathode); and RA-SVC (anode) to CS (cathode). ADER was defined as the voltage needed to defibrillate the atria in 10% to 90% of 20 consecutive shocks. Three lead systems had ADER lower than the RA (anode) to CS (cathode) configuration, which required a mean of 143+/-58 volts. These three were: RA-SVC (anode) to CS (cathode) 103+/-29 V; I-CS (anode) to RA (cathode) 129+/-39 V; and P-CS (anode) to RA (cathode) 130+/-38 V. The remaining configurations had ADER higher than the RA (anode) to CS (cathode) configuration.
Conclusion:
Adding an additional shocking electrode may reduce ADER when compared with the RA (anode) to CS (cathode) configuration. This concept could be incorporated into future implantable atrial defibrillators or used for refractory patients undergoing temporary transvenous cardioversion.
More Related Videos
12:45Benefits of Cardiac Resynchronization Therapy in an Asynchronous Heart Failure Model Induced by Left Bundle Branch Ablation and Rapid Pacing
Published on: December 11, 2017
06:57Ablation of Ischemic Ventricular Tachycardia Using a Multipolar Catheter and 3-dimensional Mapping System for High-density Electro-anatomical Reconstruction
Published on: January 31, 2019
Related Concept Videos
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin to...
Cardiac Catheterization IV: Nursing Management
Dysrhythmias VI: Management of Dysrhythmias