Pacemaker-mediated polymorphic ventricular tachycardia

D S Goldman1, P A Levine

  • 1Division of Cardiology, University of Florida, Jacksonville, USA.

Pacing and Clinical Electrophysiology : PACE
|October 30, 1998
PubMed
Summary
This summary is machine-generated.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Eye-sparing protocol in sinonasal cancers -vision at what cost?: Preservation of the eye in the treatment of sinonasal malignant neoplasms with orbital involvement.

Indian journal of otolaryngology and head and neck surgery : official publication of the Association of Otolaryngologists of India·2012
Same author

Miniaturized spectrometer employing planar waveguides and grating couplers for chemical analysis.

Applied optics·2010
Same author

Management of the patient with an acute massive rise in the capture threshold.

Indian pacing and electrophysiology journal·2006
Same author

Evaluation of omeprazole in the treatment of reflux laryngitis: a prospective, placebo-controlled, randomized, double-blind study.

The Laryngoscope·2002
Same author

Case report: complex arrhythmia induced by noncompetitive atrial pacing algorithm of DDDR pacemaker.

Journal of interventional cardiac electrophysiology : an international journal of arrhythmias and pacing·2001
Same author

The paced 12-lead electrocardiogram should no longer be neglected in pacemaker follow-up.

Pacing and clinical electrophysiology : PACE·2001
Same journal

A Case of Permanent Pacemaker Implantation via the Epicardial Approach Using the 3830 Lead in an 11-Day-Old Neonate (With Follow-Up of the Above Case).

Pacing and clinical electrophysiology : PACE·2026
Same journal

Cryoballoon Versus Radiofrequency Ablation for Persistent Atrial Fibrillation: Meta-Analysis of Randomized Trials.

Pacing and clinical electrophysiology : PACE·2026
Same journal

Tilt Test Duration in Suspected Vasovagal Syncope: Temporal Patterns and Diagnostic Yield in Patients From Central China.

Pacing and clinical electrophysiology : PACE·2026
Same journal

Combined Leadless Pacing and Subcutaneous ICD Therapy in Long QT Syndromes.

Pacing and clinical electrophysiology : PACE·2026
Same journal

Association of Anesthesia Modality With Procedural Parameters and Clinical Outcomes in PVI for Atrial Fibrillation.

Pacing and clinical electrophysiology : PACE·2026
Same journal

Zero-Fluoroscopy Cryoballoon Ablation for Paroxysmal Atrial Fibrillation in a Patient With Dextrocardia: A Case Report.

Pacing and clinical electrophysiology : PACE·2026
See all related articles

Pacemaker programming can inadvertently cause arrhythmias. Disabling a specific feature (+ PVARP on PVC) prevented recurrent polymorphic ventricular tachycardia (PMVT) in a patient with refractory bradycardia.

Area of Science:

  • Cardiology
  • Electrophysiology
  • Medical Device Technology

Background:

  • Polymorphic ventricular tachycardia (PMVT) is a serious arrhythmia often requiring intervention.
  • Permanent pacemaker implantation is a strategy for managing refractory bradycardia-dependent arrhythmias.
  • Pacemaker-mediated tachycardia (PMT) can be triggered by device programming.

Observation:

  • A patient with refractory bradycardia-dependent PMVT received a permanent pacemaker.
  • Despite pacemaker implantation, the patient experienced recurrent PMVT.
  • Device interrogation revealed a factory-programmed feature (+ PVARP on PVC) as the cause.

Findings:

  • The '+ PVARP on PVC' feature, intended to prevent PMT, induced cycle length variability.
  • This variability in cardiac cycle length precipitated recurrent PMVT episodes.

Related Experiment Videos

  • Deactivating the '+ PVARP on PVC' feature successfully resolved the arrhythmia.
  • Implications:

    • Standard pacemaker programming may require individual adjustment for certain arrhythmias.
    • Understanding device-feature interactions is crucial for preventing iatrogenic arrhythmias.
    • This case highlights the importance of thorough device interrogation and reprogramming in managing complex cardiac conditions.