Protocol of Electrochemical Test and Characterization of Aprotic Li-O2 Battery
Published on: July 12, 2016
[Evaluation of the FREEDOM O2, DC Concentrator, a portable oxygen concentrator capable of operating under car battery
T Nakamoto1, T Nakamachi, T Hase
1First Department of Internal Medicine, Dokkyo University School of Medicine, Tochigi, Japan.
Unlabelled:
The FREEDOM O2 DC Concentrator, an oxygen concentrator which can be powered by a car battery, was evaluated. The oxygen concentrator was used by a 67 year-old man with sequelae of pulmonary tuberculosis who was receiving long-term oxygen therapy at home (HOT), and whose work required automobile trips over long distances. The equipment used was an adsorption-type oxygen concentrator capable of operating on a DC 12 V power supply, and which can be powered from a residential power outlet (AC 100 V) using a dedicated voltage converter. The trunk-shaped equipment measured 21/584 x 42 cm and weighed 17 kg. The noise level of the equipment was 58.2 +/- 2.5 dB (at 1 meter), and the flow rate can be set to 0.25, 0.50, 0.75, 1.0, 1.5, and 2.0 l/min.
Results:
1) The O2 concentration which can be generated by this equipment is 93 +/- 3% (0.25 to 1.5 l/min) or 90 +/- 2.8% (2.0 l/min). 2) Using this equipment, the patient was capable of driving himself in comfort for two hours or longer. Further, it was possible to stay in a hotel during a trip, inhaling oxygen generated by the equipment. Hereafter, this equipment should enable or facilitate long-distance driving, travel and lodging for HOT patients.
Related Concept Videos
Batteries and Fuel Cells
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Respiratory Assessment: Purpose and Indications
Objectives and Importance:
The primary goal of respiratory assessment is to evaluate patients at early risk of clinical deterioration. Since respiratory distress often precedes other signs of declining health, breathing patterns and sounds become a...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Administering Oxygen by Mask
Administering oxygen by mask is a common nursing intervention that provides supplemental oxygen to patients with respiratory distress or chronic lung conditions. This procedure involves delivering oxygen at a specified rate through a face mask connected to an oxygen source.
Equipment
The equipment necessary for this procedure includes: