Characterization of the low magnification performance of a Philips CM300-FEG
1International Institute for Advanced Research, Matsushita Electric Industrial Co., Ltd., Kyoto, Japan.
Abstract:
The low magnification performance of a Philips CM300-FEG transmission electron microscope was characterized in three different configurations: CM30-FEG Super Twin, CM30-FEG Twin and CM300-FEG Cryo Twin. Micrographs of gold single crystal, polycrystalline gold, graphitized carbon and copper chloropthalocyanine were recorded in the magnification range 2050x-100,000x using film as the recording medium and optical diffraction to assay image resolution. Lattice images of gold single crystals could be recorded on film with a resolution of 2.04 A at 10,500x; copper chloropthalocyanine lattice images could be recorded with a resolution of 6 A at 4500x and 12 A at 2050x. This indicates that the film has a line resolution as low as 2.1 microns or 470 lines/mm. The fact that lattice images from different specimens can be recorded at this line spacing regardless of magnification suggests that film rather than the microscope is the major limitation to image resolution at low magnification. The image resolution delivered by the microscope is also approximately 4x smaller than the capability of densitometers to digitize it. This suggests that there would be some benefit to digitization of micrographs at finer pixel sizes than currently possible. These observations indicate that tomography of radiation sensitive specimens, where the film is the recording medium, can be recorded at the lowest magnification consistent with digitizing capabilities and desired resolution with confidence in the microscope performance. The potential benefit lies in the possibility of obtaining more images for the same accumulated dose at low magnification and the possibility that more unit cells can be imaged both of which can contribute to the potential success of applying tomographic methods to radiation sensitive specimens.
More Related Videos
10:35Multimodal Imaging and Spectroscopy Fiber-bundle Microendoscopy Platform for Non-invasive, In Vivo Tissue Analysis
Published on: October 17, 2016
08:22Measurement of 3-Dimensional cAMP Distributions in Living Cells using 4-Dimensional (x, y, z, and λ) Hyperspectral FRET Imaging and Analysis
Published on: October 27, 2020
Related Concept Videos
Imaging Biological Samples with Optical Microscopy
In optical microscopy, the specimen to be viewed is placed on a glass slide and clipped on the stage...
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...