Related Experiment Videos
Capillary ion electrophoresis screening of nerve agent degradation products in environmental samples using
1Research and Technology Directorate, US Army Edgewood Research, Development and Engineering Center, Aberdeen Proving Ground, MD 21010-5423, USA.
Abstract:
A method of detecting signature methylphosphonic acid (MPA) breakdown products of V and G nerve agents in environmental samples was developed using capillary ion electrophoresis with conductivity detection. The electrolyte (30 mM L-histidine, 30 mM 2-(N-morpholino)ethanesulfonic acid, 0.7 mM tetradecyltrimethylammonium hydroxide, and 0.03 weight% Triton X-100) allowed baseline separation of MPA, ethyl methylphosphonic acid (EMPA), isopropyl methylphosphonic acid (IMPA), and pinacolyl methylphosphonic acid (PMPA) in less than 10 min. Detector response was linear in the 6-60 micrograms/ml concentration range (correlation coefficient = 0.99) with a detection limit around 6 micrograms/ml. The application of this method for screening MPA, EMPA, IMPA, and PMPA in surface water, groundwater, and soil extracts is demonstrated.