Scanning electron microscope appearances of Lightspeed instruments used clinically: a pilot study

M Marending1, F Lutz, F Barbakow

  • 1Department of Preventive Dentistry, School of Dentistry, University of Zurich, Switzerland.

International Endodontic Journal
|November 21, 1998
PubMed
Summary

Overusing Lightspeed dental instruments can lead to microfractures in cutting heads. Clinicians should monitor instrument condition to prevent potential damage during root canal treatments.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Comparison of two contemporary rotary systems in a pre-clinical student course setting.

International endodontic journal·2015
Same author

[Posttraumatic complications after skull fracture in 24 children].

Zeitschrift fur Kinderpsychiatrie. Revue de psychiatrie infantile·2014
Same author

[An investigation with glutamic acid therapy of 16 of 30 mentally defective pupils].

Zeitschrift fur Kinderpsychiatrie. Revue de psychiatrie infantile·2014
Same author

[Psychological development of a girl aged six and a half years suffering from constitutional pubertas praecox].

Zeitschrift fur Kinderpsychiatrie. Revue de psychiatrie infantile·2014
Same author

[A case of dementia infantilis of Heller].

Zeitschrift fur Kinderpsychiatrie. Revue de psychiatrie infantile·2014
Same author

Primary assessment of a self-adhesive gutta-percha material.

International endodontic journal·2012

Area of Science:

  • Endodontics
  • Dental Materials Science

Background:

  • Lightspeed instruments are widely used in endodontic procedures.
  • Assessing the wear and tear of dental instruments is crucial for patient safety and treatment efficacy.

Purpose of the Study:

  • To evaluate the effects of clinical use on the cutting heads of Lightspeed endodontic instruments.
  • To identify potential failure modes and recommend optimal usage guidelines.

Main Methods:

  • Scanning electron microscopy (SEM) was used to examine 108 Lightspeed instrument cutting heads.
  • Instruments were sourced from two distinct clinical settings with varying usage patterns.
  • Defects such as debris, metal strips, disrupted edges, microfractures, tears, pitting, fretting, and fatigue cracks were assessed.

Related Experiment Videos

Main Results:

  • Ultrasonic cleaning reduced debris presence before SEM examination.
  • Microfractures and tears were more common in instruments used more frequently (faculty group).
  • Pitting and fretting were equally prevalent in both groups; no fatigue cracks were observed.

Conclusions:

  • Frequent use of Lightspeed instruments may increase the risk of microfractures in the cutting head flutes.
  • Clinicians should be aware of potential instrument degradation with prolonged use.
  • Adherence to recommended usage or replacement protocols is advised to maintain instrument integrity.