Related Experiment Video
Updated: Aug 4, 2026

Measuring Phagosome pH by Ratiometric Fluorescence Microscopy
Published on: December 7, 2015
Comparison of pleural fluid pH values obtained using blood gas machine, pH meter, and pH indicator strip
D S Cheng1, R M Rodriguez, J Rogers
1Department of Medicine, Saint Thomas Hospital and Vanderbilt University, Nashville, TN 37202, USA.
Study Purpose:
The purpose of this study was to compare the pleural fluid pH values obtained with a blood gas machine (pHbg), with a pH meter (pHmet), and with a pH indicator strip (pHstrip), to determine if the pleural fluid pH measured by a pH meter or a pH indicator strip was sufficiently accurate for clinical decisions.
Methods:
The pleural fluid pH was determined, within 20 min after being collected anaerobically, by a blood gas machine (CIBA-Corning model 288), pH meter (Corning pH meter 610A), and pH indicator strip (Baxter Diagnostic) following routine laboratory procedures in 50 pleural fluids. Pleural fluid pH was determined in seven additional samples with the blood gas machine and a pH meter at 25 and 37 degrees C respectively, initially, and after 30 min.
Results:
The mean pHbg (7.42+/-0.01) was significantly less than the mean pHmet (7.58+/-0.02) or the mean pHstrip (8.23+/-0.06). There were significant differences between the pHbg and the pHmet (p < 0.001), and between the pHbg and the pHstrip (p < 0.001). Analysis of the additional seven samples demonstrated that when the blood gas machine was set at 25 degrees C, the pHbg (pHbg = 7.54+/-0.02) and the pHmet (7.53+/-0.01) were almost identical.
Conclusion:
When the pleural fluid pH is going to be used for decision making, only the pH values provided by the blood gas machine are sufficiently accurate.
More Related Videos
10:18Real-Time, Semi-Automated Fluorescent Measurement of the Airway Surface Liquid pH of Primary Human Airway Epithelial Cells
Published on: June 13, 2019
04:05Clinical Application of Phase Angle and BIVA Z-Score Analyses in Patients Admitted to an Emergency Department with Acute Heart Failure
Published on: June 30, 2023
Related Concept Videos
Indicators
Pressure Gauges
Fluorescence and Phosphorescence: Instrumentation
Diagnosing Acidosis and Alkalosis
First, the pH level is assessed to determine whether the blood pH is normal (7.35–7.45), low (acidosis), or high (alkalosis).
Next, the PCO2 and HCO3− values are examined to...
Blood Studies I: ABG and VBG
Arterial Blood Gas (ABG)
Arterial Blood Gas (ABG) studies are crucial for assessing the lungs' ability to supply oxygen and remove carbon dioxide, reflecting the patient's ventilation status. They also help understand the kidneys' capacity to reabsorb or...
Measurement of Fluid Pressure
A basic form of manometer is the piezometer, a vertical tube open at the top and filled with the same...