Simple high-performance liquid chromatographic method for determination of pentoxifylline in human plasma
Related Experiment Video
Updated: Oct 27, 2026

A Convenient Method for Extraction and Analysis with High-Pressure Liquid Chromatography of Catecholamine Neurotransmitters and Their Metabolites
Published on: March 1, 2018
Jia Woei Wong1, Kah Hay Yuen, Kok Khiang Peh
1School of Pharmaceutical Sciences, University of Science Malaysia, Penang.
Abstract:
A simple high-performance liquid chromatographic method using ultraviolet detection was developed for the determination of pentoxifylline in human plasma. Prior to analysis, pentoxifylline and the internal standard (chloramphenicol) were extracted from plasma sample using dichloromethane. The mobile phase comprised 0.02 M phosphoric acid adjusted to pH 4, methanol and tetrahydrofuran (55:45:1, v/v). Analysis was run at a flow-rate of 1.4 ml/min with the detector operated at a wavelength of 273 nm. The method was specific and sensitive with a detection limit of approximately 3.0 ng/ml at a signal to noise ratio of 3:1, while the limit of quantification was 12.5 ng/ml. Mean recovery value of the extraction procedure was about 99.9%, while the within-day and between-day coefficient of variation and percent error values of the assay method were all less than 10.0%. The calibration curve was linear over a concentration range of 12.5-400.0 ng/ml.
Related Concept Videos
High-Performance Liquid Chromatography: Instrumentation
High-Performance Liquid Chromatography: Elution Process
High-Performance Liquid Chromatography: Types of Detectors