3-D sonographic volume measurement of the cerebral ventricular system: in vitro validation
W Kampmann1, M M Walka, M Vogel
1Department of Neonatology, Charité, Virchow-Hospital, Humboldt Universität Berlin, Germany. wilhelm.kampmann@t-online.de
Abstract:
This in vitro study evaluates the accuracy and observer dependency of a freehand three-dimensional (3-D) ultrasound system for measuring cerebral ventricular volume in infants. A sphere, a cylinder, and three cerebral ventricle phantoms were made of agarose and embedded in an echogenic matrix after measuring true volumes by water displacement. Volumes of the models were calculated by 3-D software after manual contour marking on ultrasound cross-sections. Mean +/- SD sonographic volume was 94.6%+/-7.3% (n = 130) of true volumes. Intraobserver variation (n = 10) was 2.3%-5.3% for complete investigation (scanning and marking), and 1.8%-4.1% for marking alone. Difference of means (n = 10) between two observers was 7.6% to 10.8% for complete investigation, and 0.6% to 1.6% for the marking process. We conclude that 3-D freehand ultrasonography may be useful for examining ventricle dilatation in infancy.