Related Experiment Video
Updated: Oct 26, 2026

A Novel Rescue Technique for Difficult Intubation and Difficult Ventilation
Published on: January 17, 2011
Continuous tracheal gas insufflation enables a volume reduction strategy in hyaline membrane disease: technical
G Dassieu1, L Brochard, E Agudze
1Service de réanimation néonatale, Hôpital Intercommunal de Créteil, France.
Objective:
Instrumental dead space wash-out can be used to improve carbon dioxide clearance. The aim of this study was to define, using a bench test, an optimal protocol for long-term use, and to assess the efficacy of this technique in neonates.
Design:
A bench test with an artificial lung model, and an observational prospective study. Dead space wash-out was performed by continuous tracheal gas insufflation (CTGI), via six capillaries molded in the wall of a specially designed endotracheal tube, in 30 preterm neonates with hyaline membrane disease.
Setting:
Neonatal intensive care unit of a regional hospital.
Results:
The bench test study showed that a CTGI flow of 0.5 l/ min had the optimal efficacy-to-side-effect ratio, resulting in a maximal or submaximal efficacy (93 to 100%) without a marked increase in tracheal and CTGI circuit pressures. In the 30 newborns, 15 min of CTGI induced a significant fall in arterial carbon dioxide tension (PaCO2), from 45 +/- 7 to 35 +/- 5 mmHg (p = 0.0001), and in 14 patients allowed a reduction in the gradient between Peack inspirating pressure and positive end-expiratory pressure from 20.8 +/- 4.6 to 14.4 +/- 3.7 cmH2O (p < 0.0001) while keeping the transcutaneous partial pressure of carbon dioxide constant. As predicted by the bench test, the decrease in PaCO2 induced by CTGI correlated well with PaCO2 values before CTGI (r = 0.58, p < 0.002) and with instrumental dead space-to-tidal volume ratio (r = 0.54, p < 0.005).
Conclusion:
CTGI may be a useful adjunct to conventional ventilation in preterm neonates with respiratory disease, enabling an increase in CO2 clearance or a reduction in ventilatory pressure.
Related Concept Videos
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Cardiopulmonary Resuscitation V: Advanced Airway Management Techniques
Trachea
Anatomical Features:
Location: About half of the trachea is situated in the neck, anterior to the esophagus, and extends from the larynx (at the level of the...
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Tracheostomy Decannulation
Description of the Procedure
Decannulation refers to the permanent removal of the tracheostomy tube, signaling the resolution of the condition that initially necessitated the tracheostomy. The process requires a well-coordinated interplay between...
Endotracheal Intubation I: Procedure
The ET tube comprises various components, including a standard adaptor to attach a bag-valve-mask (BVM) or ventilator, a cuff, a pilot balloon, and radiopaque markings along its length to measure the insertion distance. The tube sizes...