The signal in total-body plethysmography: errors due to adiabatic-isothermic difference
Related Experiment Video
Updated: Oct 26, 2026

Combining Volumetric Capnography And Barometric Plethysmography To Measure The Lung Structure-function Relationship
Published on: January 8, 2019
J G Chaui-Berlinck1, J E Bicudo
1Departamento de Fisiologia, Instituto de Biociências, Universidade de São Paulo, SP, Brazil.
Abstract:
Total-body plethysmography is a technique often employed in comparative physiology studies because it avoids excessive handling of the animals. The pressure signal obtained is generated by an increase in internal energy of the gas phase of the system. Currently, this increase in internal energy is ascribed to heating (and water vapour saturation) of the inspired gas. The standard equation for computing tidal-volume implies that only temperature and saturation differences can be responsible for generating the ventilation signal. In this study, we were able to demonstrate that the difference between the external process of the thoracic expansion, which is adiabatic, and the internal process of it, which is isothermic, is an important factor of internal energy change in the total-body plethysmography method. In other words, organic tissues transfer heat to the entering gas but also to the present gas, in a way that keeps internal expansion an isothermic process. This extra amount of energy was never taken into account before. Therefore, experiments using such a technique to measure tidal-volume should be done using isothermic chambers. Moreover, due to uncertainties of the complementary measurements (ambient and lung temperatures, ambient water vapour saturation) needed to compute tidal-volume using total-body plethysmography, a minimal temperature difference about 15 degrees C between body and ambient should exist to keep uncertainties in tidal-volume values below 5%. However, this limit is not absolute, because it varies as a function of humidity and degree of uncertainty of the complementary measurements.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Errors occurring during blood pressure monitoring
Several factors...
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...