[Clinical evaluation of a new non-radiometric automatic system for the rapid diagnosis of tuberculosis]
M Casal1, J Gutiérrez, M Vaquero
1Servicio de Microbiología Clínica, Hospital Universitario Reina Sofía, Córdoba.
Background:
Tuberculosis (TB) is again a public health problem un many countries and is considered a re-emerging disease. The fastest possible diagnosis in our patients is essential for TB control programs. ESP is a non-radioactive, totally automated, continuously monitored system designed to detect mycobacteria.
Methods:
Clinical evaluation of this new system for the rapid diagnosis of tuberculosis. During 1997 a total of 1,022 clinical sputum specimens were investigated. Specimens were processed in triplicate for ESP, BACTEC 460 TB and Löwenstein-Jensen systems. The validity, isolates of Mycobacterium tuberculosis and time required for detecting M. tuberculosis by the three systems were determined.
Results:
The sensitivity, specificity, positive predictive and negative predictive values of the new systems were 98%, 99.8%, 98% and 99.8%, respectively. No significant differences were found between the recovery rates by the three systems. The mean time for detection was 10 days (range: 7-13 days) for specimens with positive bacilloscopy and 14 days (range: 10-28 days) for specimens with negative bacilloscopy. The difference was statistically significant between ESP and Löwenstein-Jensen, but not between ESP and BACTEC.
Conclusions:
The new system proved to have an excellent sensitivity and specificity, which along with its total automation renders it a system of great clinical interest for the rapid diagnosis of TB and an alternative method for radiometric systems.