Related Experiment Videos
High-performance liquid chromatographic method for measuring total plasma homocysteine levels
I Fermo1, C Arcelloni, G Mazzola
1Laboratory of Chromatographic and Separative Techniques, IRCCS H San Raffaele, Milan, Italy. fermo.isabella@hsr.it
Journal of Chromatography. B, Biomedical Sciences and Applications
|December 30, 1998
Summary
A new high-performance liquid chromatography (HPLC) method simplifies total plasma homocysteine (tHcy) measurement using safer reagents and faster run times. This improved assay offers high reproducibility for routine clinical analysis.
Area of Science:
- Clinical Chemistry
- Analytical Chemistry
- Biochemistry
Background:
- Accurate measurement of total plasma homocysteine (tHcy) is crucial for clinical diagnostics.
- Existing methods for tHcy analysis can be complex and time-consuming.
- Optimization of analytical procedures is needed for routine clinical laboratory use.
Purpose of the Study:
- To develop and validate a modified high-performance liquid chromatographic (HPLC) method for routine clinical analysis of total plasma homocysteine (tHcy).
- To enhance the assay's efficiency, reproducibility, and safety through reagent and procedural modifications.
Main Methods:
- Modification of an existing HPLC procedure using SBD-F (ammonium-7-fluorobenzo-2-oxa-1,3-diazole-4-sulphonate) pre-column derivatization.
- Replacement of tri-n-butylphosphine with sodium borohydride as the reductant.
- Inclusion of mercaptopropionylglycine as an internal standard.
- Utilized basic isocratic HPLC equipment with a 14-minute run time.
Main Results:
- The modified HPLC assay demonstrated high reproducibility with within- and between-run coefficients of variation (C.V.) of 1.9% and 4%, respectively.
- Established reference values for fasting tHcy: 7.65±2.3 µmol/l (women) and 8.9±2.4 µmol/l (men).
- Determined post-methionine load tHcy levels: 19.9±5.5 µmol/l (women) and 26.8±5.5 µmol/l (men).
Conclusions:
- The modified HPLC method is suitable for routine clinical total plasma homocysteine analysis.
- The use of sodium borohydride and mercaptopropionylglycine improves assay safety and reproducibility.
- The optimized 14-minute run time enhances analytical efficiency for clinical laboratories.