Predictors of loss of atrioventricular synchrony in single lead VDD pacing
P Hunziker1, P Buser, M Pfisterer
1Division of Cardiology, University Hospital, Basel, Switzerland. HunzikerPA@ubaclu.unibas.ch
Insights
Maintaining proper VDD pacing function requires avoiding a low atrial dipole position. This anatomical factor, not electrical programming, predicts long-term VDD mode success in patients with atrioventricular block.
Area of Science:
- Cardiology
- Biomedical Engineering
- Medical Device Technology
Background:
- Atrioventricular (AV) block necessitates pacing solutions.
- Ventricular-Demand (VDD) pacing offers atrial synchronous ventricular pacing.
- Maintaining VDD mode function is crucial for patient outcomes.
Purpose of the Study:
- To assess the long-term maintenance of VDD pacing function.
- To identify factors predicting VDD mode failure in patients with AV block and normal sinus function.
Main Methods:
- An observational study of 86 patients with single-lead VDD pacing systems.
- Pacemaker function evaluated via event counters, telemetry, and Holter monitoring.
- Demographic, radiological, and pacing variables analyzed for correlation with VDD function loss.
Main Results:
- 81% of patients maintained proper VDD function over a mean follow-up of 10 months.
- Predictors of VDD failure included low atrial dipole position and maximum programmable atrial sensitivity.
- Multivariate analysis identified only dipole position as a significant predictor of VDD failure.
Conclusions:
- Avoiding a low atrial dipole position relative to the carina is essential for long-term VDD function.
- Electrical guidance of dipole positioning did not influence long-term VDD pacing outcomes.
Objective:
To evaluate maintenance of proper VDD function, defined as persistence of sinus rhythm with atrial synchronous ventricular pacing, and to define factors predicting failure of the VDD mode in patients with atrioventricular (AV) block and normal sinus function.
Design:
Observational study in 86 consecutive patients (mean (SD) age 74 (12) years; 38 women, 48 men) with single lead VDD pacing systems (Intermedics Unity, n = 66, Medtronic Thera VDD, n = 20), implanted for high degree AV block with documented normal sinus node. Pacemaker function was assessed by event counters, telemetric measurements, and Holter recordings. Demographic, radiological, and pacing variables were correlated with loss of proper VDD function.
Results:
During a mean (SD) follow up of 10 (10) months (range 1-37), sinus rhythm and atrial triggered ventricular pacing were maintained in 70 of 86 patients (81%). Atrial undersensing was observed in nine patients, lead migration in two, atrial fibrillation in three, and symptomatic sinus bradycardia in two. Univariate predictors of loss of proper VDD function were: low position of the atrial dipole relative to the carina (> or = 6 cm; p < 0.01) during fluoroscopy; and maximum programmable atrial sensitivity of the pacemaker (p = 0.03). In a multivariate analysis, only dipole position remained predictive of outcome (p < 0.02). Not predictive were sex, age, symptoms before pacemaker implantation, cardiothoracic ratio or dilatation of individual heart chambers on chest x ray, side of device implant, and P wave amplitude at implant.
Conclusions:
To maintain proper VDD function in the long term, a low anatomical dipole position relative to the carina should be avoided. Electrical guidance of dipole positioning does not seem to influence long term outcome.
Related Concept Videos
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...
Dysrhythmias I: Introduction
Dysrhythmias II: Classification of Tachyarrhythmias
Dysrhythmias III: Characteristics of Dysrhythmias
Dysrhythmias IV: Characteristics of Bradyarrhythmias
Dysrhythmias V: Evaluating Dysrhythmias