Determination of total and specific IgE using UNICAP 100: comparative study with the CAP system
1Allergy and Paediatric Clinical Immunology Department, Maternity and Children's University Hospital, Vall d'Hebrón, Barcelona.
Allergologia Et Immunopathologia
|January 14, 1999
Summary
The UNICAP 100 offers fully automated, reliable measurement of specific IgE, matching the CAP System
Area of Science:
- Allergy diagnostics
- Immunology
- Clinical chemistry
Background:
- Serum specific IgE tests historically had lower sensitivity than skin tests.
- The CAP system improved sensitivity but remained semi-automatic with daily calibration needs.
- The UNICAP 100 was developed as a fully automated system for IgE and Eosinophil Cationic Protein analysis.
Purpose of the Study:
- To evaluate the practicality and reliability of the UNICAP 100.
- To compare UNICAP 100 performance against the CAP System for total and specific IgE.
- To assess diagnostic sensitivity and specificity using skin tests as a reference in pediatric patients.
Main Methods:
- Comparison of UNICAP 100 with the CAP System in 150 pediatric patients.
- Assessment of total and specific IgE determination.
- Skin tests used as the gold standard for diagnostic accuracy evaluation.
Main Results:
- UNIPAC 100 demonstrated excellent interchangeability with the CAP System (correlation coefficients 0.985-0.998).
- Intra-series imprecision for total IgE was 2.1-3.6% and specific IgE was 2.2-5.1%.
- Inter-series imprecision was 3.3-7.7% for total IgE and 5.2-8.9% for specific IgE, with diagnostic sensitivity of 70-95% and specificity of 87-100%.
Conclusions:
- UNIPAC 100 provides interchangeable results with the CAP System.
- The UNICAP 100 offers significant advantages through complete automation and simplified calibration.
- This automated system enhances the practicality and reliability of specific IgE testing in allergy diagnostics.