Related Experiment Video
Updated: Aug 3, 2026

Microfluidic Flow Chambers Using Reconstituted Blood to Model Hemostasis and Platelet Transfusion In Vitro
Published on: March 19, 2016
The Xylum Clot Signature Analyzer: a dynamic flow system that simulates vascular injury
C K Li1, T J Hoffmann, P Y Hsieh
1Xylum Corporation, Scarsdale, New York 10583, USA. Conan_Li@Xylum.ccmail.compuserve.com
The Xylum Clot Signature Analyzer (CSA) assesses platelet function and blood clotting in whole blood. It measures platelet hemostasis time (PHT) and collagen-induced thrombus formation (CITF), reflecting platelet function under physiological conditions.
Area of Science:
- Hematology
- Biomedical Engineering
- Medical Diagnostics
Background:
- Assessing hemostasis requires evaluating platelet function and coagulation.
- Existing methods may not fully replicate physiological conditions.
- A user-friendly analyzer for near-physiological hemostasis assessment is needed.
Purpose of the Study:
- To evaluate the Xylum Clot Signature Analyzer (CSA) for assessing platelet function and coagulation.
- To determine if CSA parameters reflect platelet function and coagulation under physiological flow and temperature.
- To investigate the role of shear rate and specific inhibitors on CSA measurements.
Main Methods:
- The CSA analyzes non-anticoagulated whole blood in a single venipuncture.
- Measurements include platelet hemostasis time (PHT), collagen-induced thrombus formation (CITF), and clotting time (CT).
- Blood samples were tested with varying shear rates and inhibitors (anti-vWF, anti-GPIb, ATA, anti-GPIIb-IIIa, heparin).
Main Results:
- PHT and CITF increased at lower shear rates, indicating shear-dependent platelet activation.
- Inhibitors targeting von Willebrand factor (vWF) and platelet receptors prolonged PHT and CITF, especially at high shear rates.
- Heparin significantly increased clotting time (CT), demonstrating inhibition of coagulation under flow.
Conclusions:
- The CSA accurately assesses multiple critical aspects of hemostasis in whole blood.
- CSA parameters reflect platelet function and coagulation under near-physiological flow conditions.
- The analyzer is sensitive to shear-induced platelet activation and the effects of hemostatic inhibitors.