Related Experiment Videos
Determination of trace level bromate in drinking water by direct injection ion chromatography
L K Jackson1, R J Joyce, M Laikhtman
1Dionex Corporation, Sunnyvale, CA 94088-3606, USA.
Abstract:
Disinfection byproduct anions such as bromate, chlorite and chlorate pose significant health risks, even at low microgram/l levels in drinking water. A direct injection, ion chromatographic method was developed using a Dionex AS9-HC anion-exchange column with a carbonate eluent and suppressed conductivity detection for the determination of these disinfection byproduct anions, and bromide, at low microgram/l levels in drinking water. No additional sample pretreatment, other than filtration, is required. The method is linear for the oxyhalides and bromide over the typical concentration range expected for these analytes in drinking water; and quantitative recoveries were obtained for drinking water samples spiked at 10 micrograms/l. This ion chromatographic method, based on the recently developed AS9-HC column, is applicable for the quantitation of bromate in finished drinking water at the 10 micrograms/l maximum contaminant level currently being proposed by the US EPA.