Related Experiment Video
Updated: Jul 27, 2026

A Methodological Approach to Non-invasive Assessments of Vascular Function and Morphology
Published on: February 7, 2015
Interrater reliability of plaque morphology classification in patients with severe carotid artery stenosis
A Hartmann1, J P Mohr, J L Thompson
1Stroke Unit, The Neurological Institute, Columbia-Presbyterian Medical Center, New York, NY 10032, USA.
Objective:
Ultrasonographic assessment of carotid artery plaque morphology is widely used to identify patients at high risk for stroke. However, the reliability of plaque analysis in high-grade stenosis is uncertain. We determined the interrater reliability of sonographic plaque morphology analysis in patients with severe carotid artery stenosis.
Materials And Methods:
Duplex Doppler was performed on 114 patients with 80-99% stenosis of the internal carotid artery using a Siemens Quantum 2000 D with a handheld 7.5 MHz transducer. B-mode pictures with and without color coding were printed on a Sony color video printer UP-5000 W. Three raters independently evaluated plaque echolucency, heterogeneity, calcification, and surface structure. Interrater agreement was calculated by a jackknife procedure generating kappa values and two-sided 95% confidence intervals.
Results:
Kappa values and 95% confidence intervals were 0.05 (-0.07 to 0.16) for plaque surface structure, 0.15 (0.02 to 0.28) for plaque heterogeneity, 0.18 (0.09 to 0.29) for plaque echogenicity, and 0.29 (0.19 to 0.39) for plaque calcification. The upper bounds of all of the confidence intervals were below the 0.40 level suggested for minimal reliability.
Conclusion:
The low interrater agreement indicated that unaided visual assessment of static B-mode pictures to assess plaque morphology in patients with severe carotid artery stenosis is not reliable. Other evaluation procedures and standardized criteria, as yet undeveloped, are needed to improve reliability.
More Related Videos
13:45A Method to Study the Correlation Between Local Collagen Structure and Mechanical Properties of Atherosclerotic Plaque Fibrous Tissue
Published on: November 11, 2022
09:36A Magnetic Resonance Imaging-based Computational Protocol for Analysis of Plaque Morphology and Hemodynamics in Patients with Carotid Artery Stenosis
Published on: August 12, 2025
Related Concept Videos
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Atherosclerosis I: Introduction
Atherosclerosis II: Clinical Manifestations and Diagnostic Tests