Related Experiment Videos
Potentiometric sensor for methylene blue based on methylene blue-silicotungstate ion association and its
1Department of Chemistry, Yunnan University, Kunmimg, People's Republic of China.
Abstract:
A methylene blue (MB) poly(vinyl chloride) membrane sensor based on MB-silicotungstate (SLT) ion association as electroactive material was described. The linear response covered the range 1 x 10(-3)-1 x 10(-6) mol x dm(-3) MB solution, with a slope 52.0+/-0.8 mV decade(-1) (pH range 3.0-10.0). The detection limit was 7.65 x 10(-7) mol x dm(-3). The electrode showed stability, good reproducibility and fast response. Interferences from common inorganic cations, some organic base were negligible. These characteristics of the electrode enabled it to be used successfully for the determination of MB in injection. There was a good agreement for the results of MB content in injection between potentiometric method and USP standard procedure.