Related Experiment Video
Updated: Aug 10, 2026

Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
Published on: May 9, 2014
Estimating in situ exposure in the presence of acoustic nonlinearity
1Medical Physics Department, Royal United Hospital, Bath, United Kingdom.
Abstract:
Evidence is presented for excess attenuation of pulsed ultrasound due to finite amplitude effects in water. Measurements on a modern scanner are used to demonstrate that linear derating can underestimate many exposure quantities, including all safety indices apart from the cranial thermal index. More appropriate methods for estimating in situ exposure are reviewed. A preferred procedure that requires exposure measurements to be made in water under "small signal" conditions is selected. A spectral index is defined that is proposed as an indicator of finite amplitude effects, where spectral index = 0.1 defines the threshold between nonlinear and quasi-linear conditions.
Related Concept Videos
Calibration Curves: Linear Least Squares
For data that follow a straight line, the standard method for fitting is the linear...
Linear Approximation in Frequency Domain
In contrast, nonlinear systems do not inherently possess these properties. However, for small deviations around an operating point, a nonlinear system can often be approximated as linear.
Linear Approximation in Time Domain
For a simple pendulum with a mass evenly distributed along its length and the center of mass located at half the pendulum's length, the...
Application of Linearization and Approximation