Metal release and excretion from cementless titanium alloy total knee replacements
J J Jacobs1, C Silverton, N J Hallab
1Department of Orthopedic Surgery, Rush Arthritis and Orthopedics Institute, Rush Medical College, Chicago, IL 60612, USA.
Clinical Orthopaedics and Related Research
|February 12, 1999
Summary
Elevated serum titanium levels may indicate patellar component failure or wear in total knee replacements. Urine and serum levels of titanium, aluminum, and vanadium were analyzed in patients with knee arthroplasty components.
Area of Science:
- Orthopedic Surgery
- Biomaterials Science
- Toxicology
Background:
- Total knee arthroplasty (TKA) is a common procedure for knee osteoarthritis.
- Titanium alloys are frequently used in orthopedic implants due to their biocompatibility and strength.
- Wear debris and ion release from implants are potential concerns.
Purpose of the Study:
- To investigate serum and urine concentrations of titanium, aluminum, and vanadium in patients with different types of cementless primary total knee arthroplasty components.
- To assess if metal ion levels correlate with component design, wear, or failure.
Main Methods:
- Serum and urine samples were collected from patients with various TKA component designs (carbon fiber reinforced polyethylene, ultrahigh molecular weight polyethylene, nitrogen ion-implanted surfaces) and from a control group.
- Patients with failed patellar components were also included.
- Concentrations of titanium, aluminum, and vanadium were measured using appropriate analytical techniques.
Main Results:
- Serum titanium concentrations were significantly higher (approximately 50-fold) in patients with failed patellar components and (approximately 10-fold) in those with carbon fiber reinforced polyethylene bearing surfaces compared to control and other implant groups.
- No significant differences in serum or urine concentrations of aluminum or vanadium were observed among any groups.
- Urine analysis showed no significant differences in titanium, aluminum, or vanadium levels across all patient groups.
Conclusions:
- Elevated serum titanium may serve as a biomarker for patellar component failure or accelerated wear in titanium alloy-bearing total knee replacements.
- The toxicological implications of these elevated serum titanium levels remain to be determined.