Performance characteristics of the PolyTiter Immunofluorescent Titration system for determination of antinuclear

Karen A Flessland1, Helen R Landicho, Kimberlee K Borden

  • 1Alidex, Inc., Redmond, Washington 98052, USA. kflessland@alidexinc.com

Insights

The PolyTiter system offers a reliable method for quantifying antinuclear antibodies (ANA) in immunofluorescent assays (IFA). This system provides accurate endpoint titers directly from a 1:40 dilution, simplifying routine clinical laboratory testing.

Area of Science:

  • Immunology
  • Clinical Diagnostics
  • Laboratory Medicine

Background:

  • Conventional antinuclear antibody (ANA) screening uses immunofluorescent (IF) microscopy with serial dilutions.
  • Quantifying IF staining intensity traditionally involves twofold serial dilutions to determine endpoint titers.
  • This process can be time-consuming and subject to variability.

Purpose of the Study:

  • To evaluate the performance characteristics of the PolyTiter Immunofluorescent Titration system for ANA-IF assay (IFA) testing.
  • To compare the PolyTiter system's accuracy and precision against conventional serial dilution methods.
  • To assess the system's suitability for routine use in clinical laboratories.

Main Methods:

  • The PolyTiter system relates IF staining intensity to reference calibrators for direct endpoint titer determination from a 1:40 dilution.
  • Two technologists performed replicate evaluations of specimens and controls over five days.
  • Performance was assessed by endpoint dilution agreement, precision analysis (coefficients of variation), and lot-to-lot variation.

Main Results:

  • Endpoint dilution agreement with the reference was 100% for controls and high-specimen samples by one technologist, and 98-100% for others.
  • Between-reader, between-run, and within-run precision coefficients of variation were below 10%.
  • Overall agreement between PolyTiter and serial dilution for 125 specimens was 86-93%, with similar pattern identification.

Conclusions:

  • The PolyTiter system demonstrates acceptable performance for routine ANA-IFA testing.
  • It offers a reliable and potentially more efficient alternative to traditional serial dilution methods.
  • The system provides accurate and reproducible results for ANA antibody quantification.