Microfabricated on-chip-type electrochemical flow immunoassay system for the detection of histamine released in whole
Tae-Kyu Lim1, Hiroko Ohta, Tadashi Matsunaga
1Department of Biotechnology and Life Sciences, Tokyo University of Agriculture and Technology, Koganei, Tokyo 184-8588, Japan.
Insights
This study presents a novel on-chip electrochemical immunoassay for rapid histamine detection in whole blood. The system offers a fast, sensitive, and reproducible method for analyzing histamine levels, correlating well with radioimmunoassay.
Area of Science:
- Electrochemistry
- Analytical Chemistry
- Biotechnology
Background:
- Histamine is a key mediator in allergic reactions and physiological processes.
- Accurate and rapid measurement of histamine is crucial for clinical diagnostics.
- Existing methods for histamine detection can be time-consuming or require large sample volumes.
Purpose of the Study:
- To develop an on-chip electrochemical flow immunoassay system for histamine detection.
- To utilize a multichanneled matrix column for protein separation and analysis.
- To establish a rapid, sensitive, and reproducible assay for histamine in whole blood.
Main Methods:
- An on-chip electrochemical flow system with a cation-exchange resin-coated matrix column was designed.
- Ferrocenemonocarboxylic acid (Fc)-conjugated immunoglobulin G (IgG) antibody was used as a detection reagent.
- Separation of antibody-antigen complexes from free Fc-conjugated IgG was achieved based on isoelectric point (pI).
- Electrochemical detection was employed to quantify histamine concentration.
Main Results:
- The assay demonstrated a linear relationship between current and histamine concentration from 200-2000 ng/mL.
- Histamine in whole blood samples was successfully assayed within 2 minutes.
- The developed electrochemical immunoassay showed good correlation with conventional radioimmunoassay (RIA).
- The system requires only minute blood samples and yields highly reproducible results.
Conclusions:
- The on-chip electrochemical flow immunoassay is a rapid and sensitive method for histamine detection.
- This technology offers a viable alternative to conventional methods, requiring less sample and time.
- The system's performance suggests its potential for point-of-care diagnostic applications.
Abstract:
This paper describes an on-chip-type electrochemical flow immunoassay system with a multichanneled matrix column. The multichanneled matrix column was functionally coated with cation-exchange resin and used for separation of proteins. Antihistamine immunoglobulin G (IgG) antibody conjugated with ferrocenemonocarboxylic acid (Fc) was also prepared and used as a novel analytical reagent. Antibody-antigen complexes were separated from free Fc-conjugated IgG antibody (Fc-IgG) on the basis of differences in isoelectric point (pI) using the multichanneled matrix column coated with cation-exchange resin. The assay yields a good relationship between current and histamine concentration in the range of 200-2000 ng/mL. This simple technique enables the assay of histamine released in whole blood within 2 min. Furthermore, a good correlation was found between the response of the electrochemical immunoassay described in this paper and the conventional RIA (radioimmunoassay). This on-chip-type electrochemical flow immunoassay requires only minute quantities of whole blood samples and generates highly reproducible results.
More Related Videos
13:42Dry Film Photoresist-based Electrochemical Microfluidic Biosensor Platform: Device Fabrication, On-chip Assay Preparation, and System Operation
Published on: September 19, 2017
07:10Application of Biochip Microfluidic Technology to Detect Serum Allergen-specific Immunoglobulin E (sIgE)
Published on: April 21, 2019