Related Experiment Video
Updated: Aug 29, 2026

The Cultivation, Growth, and Viability of Lactic Acid Bacteria: A Quality Control Perspective
Published on: June 16, 2022
Exploring the links between quality assurance and laboratory resources. An audit-based study
Navjeevan Singh1, Aru Panwar, Vipin Fazal Masih
1Cytopathology Section, Department of Pathology, University College of Medical Sciences, Guru Tegh Bahadur Hospital, Delhi, India.
Insights
Implementing corrective measures significantly improved Ziehl-Neelsen (Z-N) staining for acid-fast bacilli (AFB) detection in cytology. This quality assurance initiative enhanced diagnostic accuracy for tuberculosis, benefiting laboratory workflow and personnel performance.
Area of Science:
- Medical Laboratory Science
- Microbiology
- Cytopathology
Background:
- Ziehl-Neelsen (Z-N) staining is crucial for diagnosing tuberculosis (TB) in cytology.
- Quality assurance is essential to maintain the accuracy of Z-N staining.
- Previous laboratory practices may have led to suboptimal diagnostic yields.
Purpose of the Study:
- To identify and resolve issues affecting Ziehl-Neelsen (Z-N) staining quality.
- To improve the accuracy of acid-fast bacillus (AFB) detection in fine needle aspiration cytology.
- To enhance overall laboratory quality assurance for TB diagnostics.
Main Methods:
- An audit-based quality assurance study was conducted on 1,421 patients with suspected tubercular lymphadenopathy.
- Laboratory practices for Z-N smear selection were evaluated over an 8-month period (Group 1).
- A two-step corrective intervention was implemented and assessed over 2 months (Group 2), with statistical analysis using the chi-squared test.
Main Results:
- In Group 1 (n=1,172), 368 diagnoses were not tuberculosis, and overall AFB positivity was 42%.
- Following corrective measures in Group 2 (n=249), AFB positivity dramatically increased to 89%.
- The observed improvement in AFB positivity was statistically significant (P < .0001).
Conclusions:
- Laboratory quality assurance issues directly impact the effectiveness of Z-N staining.
- Addressing routine laboratory problems yields substantial benefits for personnel, resource management, and workflow efficiency.
- The implemented corrective measures significantly enhanced the diagnostic yield of AFB detection through Z-N staining.
Objective:
To investigate and rectify the problems related to Ziehl-Neelsen (Z-N) staining in a cytology laboratory in the context of quality assurance.
Study Design:
An audit based quality assurance study of 1,421 patients with clinical diagnoses of tubercular lymphadenopathy who underwent fine needle aspiration cytology. Data from 8 months were audited (group 1). Laboratory practices related to selection of smears for Z-N staining were studied. A 2-step corrective measure based on results of the audit was introduced for 2 months (group 2). Results were subjected to statistical analysis using the chi 2 test.
Results:
Of 1,172 patients in group 1,368 had diagnoses other than tuberculosis. Overall acid-fast bacillus (AFB) positivity was 42%. AFB positivity in 249 patients in group 2 was 89% (P < .0001).
Conclusion:
Several issues in the laboratory are linked to quality assurance. Solving everyday problems can have far-reaching benefits for the performance of laboratory personnel, resources and work flow.
Related Concept Videos
Quality Assurance
Quality Control
Quality control helps track data, visualize trends, and identify variations, making it easier to detect deviations that may affect the accuracy of an analysis. One way to do this is by generating a quality control chart, which...
Data Validation
Key parameters for method validation include:
Types of Biopharmaceutical Studies: Controlled and Non-Controlled Approaches
Non-controlled studies, commonly employed for initial exploration, lack a control group, rendering them susceptible to biases and external influences. In contrast, controlled...
Introduction to Statistical Process Control
Systematic Error: Methodological and Sampling Errors
Sampling errors originate from improper sampling methods or the wrong sample population. These errors can be minimized by refining the sampling strategy. Defective instruments or faulty calibrations are the sources of instrumental...