Identification and quantification of high fluorescence-stained lymphocytes as antibody synthesizing/secreting cells

J Linssen1, V Jennissen, J Hildmann

  • 1Sysmex Europe GmbH, 22848 Norderstedt, Germany. Linssen.jo@sysmex-europe.com

Insights

The SYSMEX XE-2100 high fluorescence lymphocytes count accurately identifies antibody-synthesizing cells (ASC) in reactive diseases. This automated method offers a reliable tool for screening and monitoring patients with suspected infections.

Area of Science:

  • Hematology
  • Immunology
  • Clinical Chemistry

Background:

  • The SYSMEX XE-2100 offers a high fluorescence lymphocytes (HFL) count in its leucocyte differential channel.
  • Identifying HFL cells as antibody-synthesizing or -secreting cells (ASC) in reactive diseases is crucial for diagnosis.

Purpose of the Study:

  • To classify and quantify HFL-count as ASC in reactive diseases.
  • To evaluate HFL-count's accuracy against other cell populations and diagnostic methods.

Main Methods:

  • Analysis of 85 patients using the SYSMEX XE-2100.
  • Comparison with automated image analysis (Cellavision Diffmaster 96) and immunophenotyping (BD FACSCalibur).

Main Results:

  • HFL-count demonstrated high reproducibility and linearity (R(2) = 0.99).
  • Excellent correlation found between HFL-count and ASC identified by peripheral blood smear (R(2) = 0.900) and immunophenotyping (R(2) = 0.897).
  • No significant correlation was observed with monocytes or immature granulocytes.

Conclusions:

  • The SYSMEX XE-2100 HFL-count precisely identifies and quantifies ASC (activated B-lymphocytes).
  • This automated count serves as a potential screening and monitoring tool for suspected infections.
  • Further research is needed to explore the full clinical utility of HFL (ASC) counts in diagnosing inflammation, infection, or sepsis.
Abstract