Effect of serum content and diluent selection on assay sensitivity and signal intensity in multiplex bead-based

Christian Pfleger1, Nanette Schloot, Frank ter Veld

  • 1Institute for Clinical Diabetes Research, German Diabetes Center, Leibniz Institute at the Heinrich-Heine-University, Auf'm Hennekamp 65, D-40225 Duesseldorf, Germany. ch.pfleger@web.de

Insights

Choosing the right diluent is crucial for Luminex multiplex immunoassays. RD6 demonstrated superior performance over DY997, showing less signal reduction with increased human serum concentrations in cytokine determination.

Area of Science:

  • Biochemistry
  • Immunology
  • Analytical Chemistry

Background:

  • Complex biological samples like human serum require dilution in multiplexed immunoassays to mitigate matrix effects.
  • Diluents must mimic the sample matrix for accurate interpolation and ensure optimal binding for all target analytes.
  • The choice of diluent significantly impacts multiplex immunoassay results, necessitating quality assessment methods.

Purpose of the Study:

  • To evaluate and compare the performance of two commercial diluents (DY997 and RD6) in a Luminex 19-plex immunoassay.
  • To assess the impact of human serum concentration on signal intensity and assay sensitivity using different diluents.
  • To determine the most appropriate diluent for cytokine analysis in human serum samples using the Luminex system.

Main Methods:

  • A 19-plex immunoassay was performed using the Luminex system.
  • Two commercial diluents, DY997 and RD6, were tested.
  • Spiked samples with varying concentrations of human serum (25% and 50% v/v) and protein were analyzed.

Main Results:

  • Diluent DY997 showed a 55% reduction in signal intensity with 50% human serum compared to 25% serum.
  • Diluent RD6 exhibited only a 20% signal intensity reduction with 50% human serum compared to 25% serum.
  • DY997 demonstrated decreased assay sensitivity with increasing protein concentrations, more so than RD6.

Conclusions:

  • Diluent RD6 is more suitable than DY997 for cytokine determination in human serum using the Luminex multiplexed immunoassay system.
  • The study highlights the importance of diluent selection for maintaining signal intensity and assay sensitivity in complex matrices.
  • Performance metrics including signal intensity, assay sensitivity, and background levels are influenced by serum volume-fraction and protein concentration.