Evaluation of enzyme-linked immunosorbent assays for detecting circulating antibodies to Candida albicans
Harry E Prince1, Cindy Yeh, Navid Alem
1Focus Diagnostics, Inc. Cypress, California, USA.
Insights
This study validates a new Candida Detect ELISA system using cytoplasmic antigens for detecting systemic candidiasis. The assay shows excellent performance and reproducibility for IgG, IgM, and IgA antibodies, proving useful in clinical settings.
Area of Science:
- Medical Diagnostics
- Immunology
- Microbiology
Background:
- Systemic candidiasis diagnosis benefits from Candida albicans antibody assays.
- Cytoplasmic antigens are more effective than cell wall components for these assays.
Purpose of the Study:
- To evaluate the performance of a commercial Candida Detect ELISA system.
- To assess the reliability and reproducibility of measuring C. albicans IgG, IgM, and IgA antibodies.
Main Methods:
- Assessed intra-assay and inter-assay variation.
- Evaluated specimen storage stability and inter-laboratory reproducibility.
- Determined seroprevalence rates in different serum panels.
Main Results:
- Assay showed low variation (<5% intra-assay, <10% inter-assay) and good linearity.
- Excellent inter-laboratory reproducibility (>93% concordance).
- Seroprevalence rates for IgG, IgM, and IgA were established, with similar rates in control and cross-reactive panels.
Conclusions:
- The Candida Detect ELISA system demonstrates acceptable performance characteristics.
- The assay is reliable for measuring C. albicans specific antibodies.
- This system is a valuable tool for diagnosing systemic candidiasis.
Abstract:
Published studies indicate that Candida albicans antibody assays utilizing cytoplasmic antigens offer greater utility for identifying cases of systemic candidiasis when compared with assays utilizing cell wall components. We assessed the performance characteristics of a commercially available system that utilizes cytoplasmic antigens to measure C. albicans IgG, IgM, and IgA (Candida Detect ELISA reagents). Intra-assay variation was < or =5%, inter-assay variation was < or =10%, and good linearity was observed for all the three antibody isotypes. Results for specimens stored under various conditions were comparable to those obtained initially. Inter-laboratory reproducibility was excellent; qualitative concordance was > or =93% for all the three isotypes, with slopes and R(2) values approaching 1.0 in linear regression analyses. Seroprevalence in persons without apparent systemic candidiasis was evaluated using three different serum panels; seroprevalence rates ranged from 24 to 32% for IgG, 2-14% for IgM, and 15-36% for IgA. Seroprevalence rates in a panel of sera containing antibodies to other fungi were similar to rates observed in panels from individuals without systemic candidiasis. These findings demonstrate the acceptable performance of assay systems employing Candida Detect ELISA reagents.