Related Experiment Videos
Immunoglobulin-associated creatine kinase masquerading as macro-creatine kinase type 2 in a statin user
Tze Ping Loh1, Yan Hoon Ang, Siew Fong Neo
1Department of Laboratory Medicine, National University Hospital, Singapore. lohtp@yahoo.com
Insights
Macro-creatine kinase (CK) can cause falsely elevated CK levels. This case highlights how immunoglobulin-associated macro-CK type 1 can mimic macro-CK type 2, emphasizing the need for careful laboratory investigation.
Area of Science:
- Clinical chemistry
- Immunology
- Biochemical diagnostics
Background:
- Macro-creatine kinase (CK) is a known cause of artifactually elevated total CK levels.
- Macro-CK type 1 is typically immunoglobulin-associated, while type 2 is polymeric mitochondrial CK.
Observation:
- An asymptomatic elderly female patient presented with elevated CK levels following statin therapy.
- CK gel electrophoresis revealed coexisting macro-CK type 1 and type 2 patterns.
Findings:
- Further immunofixation and serum mixing studies identified an IgG-associated macro-CK (type 1) mimicking the electrophoretic pattern of macro-CK type 2.
- This unusual case demonstrates a potential for misinterpreting macro-CK type 1 as type 2.
Implications:
- Accurate differentiation of macro-CK subtypes is crucial for correct clinical interpretation.
- Falsely elevated CK levels necessitate thorough laboratory investigation, even with advanced instrumentation.
Abstract:
Macro-creatine kinase (CK) is a cause of falsely elevated CK. Macro-CK type 1 is immunoglobulin-associated CK; type 2 is polymeric mitochondrial-CK. An elderly asymptomatic lady had an elevated CK level after receiving statin therapy. Her CK gel electrophoresis analysis demonstrated coexisting macro-CK type 1 and type 2 patterns. Further analysis by immunofixation and mixing this patient's serum with CK control material revealed an IgG-associated macro-CK that mimicked the electrophoretic pattern of macro-CK type 2. This highly unusual discovery suggests the possibility of the misinterpretation of macro-CK type 1 as macro-CK type 2. Falsely elevated CK is still common despite modern laboratory instrumentation and should be investigated.
Related Concept Videos
Blood Studies for Cardiovascular System I: Cardiac Biomarkers
The essential diagnostic tools for detecting myocardial necrosis and monitoring individuals suspected of having acute coronary syndrome (ACS) include:
Troponins
Troponins, particularly cardiac troponins I and T, are the most precise and sensitive markers of myocardial injury. They are detectable within 4-6 hours of myocardial injury and remain...
Myasthenia Gravis ll: Pathophysiology
Myocarditis II: Clinical Features and Diagnostic Tests
Acute Coronary Syndrome I: Introduction
Lipid-Lowering Drugs: Statins and Miscellaneous Agents
Calmodulin-dependent Signaling
The Ca2+-CaM complex does not have enzymatic activity by itself. Instead, the complex binds downstream target proteins, including membrane proteins or enzymes,...