Report of a trial with lyophilised QC sera in the CMCH/ACBI EQAS programmes in India

R Selvakumar1, S Swaminathan, J J Fleming

  • 1Department of Clinical Biochemistry, Christian Medical College and Hospital, Vellore, Tamil Nadu.

Insights

Lyophilized quality control (QC) materials are suitable for laboratory use, showing no performance deterioration compared to liquid QC. This study confirms good stability and reconstitution techniques for lyophilized QC in clinical chemistry.

Area of Science:

  • Clinical Chemistry
  • Laboratory Medicine
  • Quality Control

Background:

  • Lyophilized quality control (QC) materials were introduced to supplement existing liquid QC programs.
  • The study involved 240 laboratories over six months, assessing 18 analytes, including new additions like LDH and triglycerides.
  • Ion-selective electrode (ISE) methods were used for sodium and potassium measurements in QC for the first time.

Purpose of the Study:

  • To compare laboratory performance using lyophilized QC versus liquid QC.
  • To evaluate the stability and effectiveness of lyophilized QC materials.
  • To identify potential standardization issues with specific analytes and methods.

Main Methods:

  • Statistical comparison of methodwise mean variance index score (MVIS) values.
  • Analysis of performance for 18 analytes using lyophilized QC materials.
  • Evaluation of laboratory ranking patterns based on overall MVIS (OMVIS).

Main Results:

  • Laboratory performance was similar for five analytes (glucose, albumin, cholesterol, SGOT, SGPT) between liquid and lyophilized QC.
  • Performance was significantly better for urea, calcium, and creatinine with lyophilized QC.
  • Significant differences were noted for total protein, sodium, potassium, and ALP, with a negative bias for sodium and potassium ISE methods.

Conclusions:

  • Introduction of lyophilized QC does not degrade laboratory performance.
  • Lyophilized QC materials exhibit good stability and support effective reconstitution techniques.
  • The study identified specific areas for improvement, including standardization of LDH and amylase reporting and sodium/potassium ISE methods.