Application of Biochip Microfluidic Technology to Detect Serum Allergen-specific Immunoglobulin E sIgE
Published on: April 21, 2019
Is the determination of specific IgE against components using ISAC 112 a reproducible technique?
Rubén Martínez-Aranguren1, María T Lizaso2, María J Goikoetxea1
1Department of Allergology and Clinical Immunology, Clínica Universidad de Navarra, Pamplona, Spain.
Insights
The ImmunoCAP ISAC 112 assay reliably detects specific IgE to 112 allergens. This fluoro-immunoassay demonstrates good to very good repeatability and reproducibility across different assays, batches, and laboratories for allergy diagnosis.
Area of Science:
- Allergen immunotherapy
- Molecular diagnostics
- In vitro diagnostics
Background:
- The ImmunoCAP ISAC 112 is a fluoro-immunoassay for detecting specific IgE to 112 molecular components from 51 allergenic sources.
- Assessing the reliability of diagnostic tools is crucial for accurate clinical decision-making in allergy management.
Purpose of the Study:
- To evaluate the intra- and inter-assay reliability of the ImmunoCAP ISAC 112.
- To assess the inter-batch and inter-laboratory reproducibility of the ImmunoCAP ISAC 112 assay.
Main Methods:
- Twenty samples, including a manufacturer calibrator (CTR02) and sera from 19 polysensitized allergic patients, were analyzed.
- Intra- and inter-assay variability was assessed by repeating measurements ten times.
- Inter-batch and inter-laboratory reproducibility were evaluated using two kit batches and two testing laboratories.
Main Results:
- The CTR02 calibrator showed significant differences from expected values in 8 of 15 allergens.
- Intra- and inter-assay agreement was very good for 94 allergens.
- Inter-batch agreement was very good for 82 allergens, and inter-laboratory agreement was very good for 73 allergens.
Conclusions:
- The ImmunoCAP ISAC 112 is a reliable in vitro diagnostic tool for specific IgE determination.
- The assay demonstrates good to very good repeatability and reproducibility, supporting its use beyond a single laboratory setting.
Background:
The ImmunoCAP ISAC 112 is a fluoro-immunoassay that allows detection of specific IgE to 112 molecular components from 51 allergenic sources. We studied the reliability of this technique intra- and inter- assay, as well as inter-batch- and inter-laboratory-assay.
Methods:
Twenty samples were studied, nineteen sera from polysensitized allergic patients, and the technique calibrator provided by the manufacturer (CTR02). We measured the sIgE from CTR02 and three patients' sera ten times in the same and in different assays. Furthermore, all samples were tested in two laboratories and with two batches of ISAC kit. To evaluate the accuracy of ISAC 112, we contrasted the determinations of CTR02 calibrator with their expected values by T Student test. To analyse the precision, we calculated the coefficient of variation (CV) of the 15 allergens that generate the calibration curve, and to analyse the repeatability and the reproducibility, we calculated the intraclass coefficient correlation (ICC) to each allergen.
Results:
The results obtained for CTR02 were similar to those expected in 7 of 15 allergens that generate the calibration curve, whereas in 8 allergens the results showed significant differences. The mean CV obtained in the CTR02 determinations was of 9.4%, and the variability of sera from patients was of 22.9%. The agreement in the intra- and inter-assay analysis was very good to 94 allergens and good to one. In the inter-batch analyse, we obtained a very good agreement to 82 allergens, good to 14, moderate to 5 allergens, poor to one, and bad to 1 allergen. In the inter-laboratory analyse, we obtained a very good agreement to 73 allergens, good to 22, moderate to 6 and poor to two allergens.
Conclusion:
The allergen microarray immunoassay, ISAC 112, is a repeatable and reproducible in vitro diagnostic tool for determination of sIgE beyond the own laboratory.
More Related Videos
06:34A Component-resolved Diagnostic Approach for a Study on Grass Pollen Allergens in Chinese Southerners with Allergic Rhinitis and/or Asthma
Published on: June 4, 2017
06:15Characterization of Thymus-dependent and Thymus-independent Immunoglobulin Isotype Responses in Mice Using Enzyme-linked Immunosorbent Assay
Published on: September 7, 2018
Related Concept Videos
Immunocytochemistry and Immunohistochemistry
These...
Enzyme-Linked Immunosorbent Assay
There are many different types of ELISAs, but they all involve an antibody molecule whose constant region binds an enzyme, leaving the variable region free to bind its specific antigen. Enzyme-substrate reaction allows the antigen to be visualized or...