Comparison of chemiluminescent immunoassay and ELISA for measles IgG and IgM

Fernando de Ory1,2, Teodora Minguito1, Pilar Balfagón1

  • 1Laboratorio de Serología, Centro Nacional de Microbiología, Instituto de Salud Carlos III, Majadahonda, Madrid, Spain.

Insights

A new automated assay accurately detects measles IgG and IgM antibodies, crucial for measles elimination efforts. This chemiluminescent immunoassay (CLIA) offers high sensitivity and specificity, simplifying laboratory diagnosis.

Area of Science:

  • Immunology
  • Virology
  • Clinical Diagnostics

Background:

  • Accurate identification of recent measles infections is vital for global measles elimination strategies.
  • Serological diagnosis relies on detecting specific immunoglobulin G (IgG) and immunoglobulin M (IgM) antibodies.
  • Automated assays streamline laboratory workflows for infectious disease diagnostics.

Purpose of the Study:

  • To evaluate the performance of an automated chemiluminescent immunoassay (CLIA) for quantifying measles-specific IgG and IgM antibodies.
  • To compare the diagnostic accuracy of the CLIA assay against a reference Enzygnost ELISA method.
  • To assess the clinical utility of the CLIA assay in identifying recent measles infections.

Main Methods:

  • Analyzed 204 sera for measles IgM and 162 sera for measles IgG using the Liaison CLIA assay.
  • Compared CLIA results with Enzygnost ELISA, with discrepancies resolved by indirect immunofluorescence.
  • Calculated sensitivity and specificity for both IgM and IgG detection.

Main Results:

  • For measles IgM, CLIA demonstrated high sensitivity (94%) and specificity (100%) against ELISA, corrected to 100% and 99.4% respectively after discrepancy resolution.
  • For measles IgG, CLIA showed excellent sensitivity (97.2%) and specificity (92.9%) against ELISA, corrected to 95.5% and 100% respectively.
  • The CLIA assay effectively detected measles IgG and IgM antibodies without the need for specimen pre-aliquoting.

Conclusions:

  • The automated CLIA assay exhibits excellent sensitivity and specificity for detecting measles IgG and IgM antibodies.
  • This assay simplifies measles serological diagnosis, eliminating pre-analytical steps like specimen aliquotting.
  • The CLIA assay is a reliable tool for clinical laboratories supporting measles elimination programs.