Heterophile antibody interference in qualitative urine/serum hCG devices: Case report

Khushbu K Patel1, Ann M Gronowski1

  • 1Department of Pathology and Immunology, Washington University School of Medicine, St. Louis, MO, USA.

Clinical Biochemistry
|March 13, 2016
PubMed

Insights

A false positive serum human chorionic gonadotropin (hCG) test in a leukemia patient was caused by heterophile antibodies. Confirmation with quantitative hCG testing is crucial when clinical suspicion arises.

Area of Science:

  • Clinical Chemistry
  • Immunodiagnostics
  • Point-of-Care Testing

Background:

  • Heterophile antibodies can interfere with immunoassays, leading to inaccurate results.
  • A case of false positive serum human chorionic gonadotropin (hCG) was observed in a patient with chronic myeloid leukemia.

Observation:

  • A qualitative hCG test yielded a positive result, but a quantitative test was negative.
  • Pretreatment with heterophilic blocking reagent (HBR) and testing on different qualitative hCG devices revealed interference.
  • Blocking reagent from an automated quantitative immunoassay successfully inhibited heterophile antibody interference.

Findings:

  • Heterophile antibody interference was identified as the cause of the false positive qualitative hCG result.
  • Different qualitative hCG devices showed varied responses due to differing antibody sources.
  • The blocking reagent from a quantitative assay effectively neutralized the interfering antibodies.

Implications:

  • Qualitative point-of-care hCG tests require cautious interpretation.
  • Quantitative hCG assays are essential for confirmation when clinical suspicion is present.
  • Understanding immunoassay interference is vital for accurate diagnostic testing.