Manual interferometric device for routine non-invasive tear film break-up time assessment

Sania Vidas Pauk1, Igor Petriček1,2, Martina Tomić3

  • 1Department of Ophthalmology, Zagreb University Hospital Center, Zagreb, Croatia.

Insights

Non-invasive tear break-up time (NIBUT) measured with a handheld device is a reliable and objective method for diagnosing dry eye disease. This new NIBUT test shows significant differences between dry eye patients and controls, outperforming standard tests.

Area of Science:

  • Ophthalmology
  • Optometry
  • Tear Film Dynamics

Background:

  • Dry eye disease (DED) affects a significant portion of the population, impacting quality of life.
  • Accurate and objective diagnostic methods are crucial for effective DED management.
  • Current diagnostic tools, including standard tear break-up time (TBUT), have limitations.

Purpose of the Study:

  • To evaluate the feasibility of non-invasive tear break-up time (NIBUT) assessment using a handheld instrument.
  • To compare the efficacy of handheld NIBUT with standard TBUT and other dry eye diagnostic tests.
  • To determine the diagnostic accuracy and predictive value of NIBUT in identifying dry eye symptoms.

Main Methods:

  • 108 subjects (56 with DED, 52 controls) were enrolled.
  • Dry eye severity was assessed using the Schein questionnaire.
  • Ocular surface parameters measured included NIBUT, TBUT, lipid layer thickness (LLT), lid-parallel conjunctival folds (LIPCOF), conjunctival hyperemia, and corneal staining.

Main Results:

  • NIBUT showed the most significant difference between dry eye patients and controls (7s vs. 17.5s, p < .001).
  • NIBUT correlated significantly with TBUT, LLT, LIPCOF, conjunctival hyperemia, and corneal staining (p < .019).
  • Short NIBUT values were superior predictors of dry eye complaints compared to TBUT and other tests (AOR = 0.87, p < .001).

Conclusions:

  • Handheld NIBUT is a simple, accessible, practical, reproducible, and objective method for assessing tear film stability.
  • NIBUT assessment holds potential for regular clinical use in diagnosing and monitoring dry eye disease.
  • This non-invasive technique offers a valuable alternative or adjunct to existing dry eye diagnostic methods.