External quality assessment of flow cytometric bronchoalveolar lavage cellular analysis: 20 years' experience in The

A H Leontine Mulder1, Harrie H M Eidhof2, Jan W Gratama3

  • 1Medlon B.V., Department of Clinical Chemistry and Laboratory Medicine, Enschede and Clinical Chemistry, Ziekenhuis Groep Twente, Almelo, The Netherlands.

Insights

A dedicated external quality assessment (EQA) for bronchoalveolar lavage (BAL) fluid analysis was developed. This EQA ensures reliable cellular analysis for interstitial lung disease diagnosis, achieving adequate quality control.

Area of Science:

  • Immunology
  • Pulmonary Medicine
  • Clinical Chemistry

Background:

  • Bronchoalveolar lavage (BAL) cellular analysis aids interstitial lung disease diagnosis.
  • Conventional flow cytometry of BAL cells is challenging due to cell properties.
  • Existing external quality schemes (EQA) for blood are not representative for BAL fluid analysis.

Purpose of the Study:

  • To establish and evaluate a dedicated EQA for BAL fluid cellular analysis.
  • To assess the quality of flow cytometric analysis of BAL cells over 20 years.
  • To determine the representativeness of BAL fluid analysis compared to peripheral blood.

Main Methods:

  • A specialized EQA procedure for stabilized BAL cells was implemented.
  • Flow cytometric parameters (lymphocyte subsets, eosinophils, neutrophils) were recorded from 2000-2020.
  • Data analysis included calculating means, standard deviations, and comparing results over time and by disease.

Main Results:

  • Median standard deviation was below 10% for most parameters, except TLY and CD103+CD3+ lymphocytes.
  • No significant time-based improvement was observed for most cell populations, except CD3+CD4+ lymphocytes.
  • Interpretation consensus varied by disease, with highest agreement for sarcoidosis and lowest for nonspecific interstitial lung disease.

Conclusions:

  • A dedicated EQA for BAL fluid cellular analysis is justified due to its unique characteristics.
  • Adequate analytical and post-analytical quality control can be achieved with this specialized EQA.
  • This EQA improves the reliability of BAL fluid analysis for interstitial lung disease diagnosis.