Serum cortisol assay performance following the 1 mg overnight dexamethasone suppression test

Jonathan S Atkins1, James M Hawley1, Laura J Owen2,3

  • 1Department of Clinical Biochemistry, Wythenshawe Hospital, Manchester University NHS Foundation Trust, Manchester, UK.

Insights

The 1 mg overnight dexamethasone suppression test (ONDST) performance varies by immunoassay platform. Roche and Siemens assays showed closer alignment with LC-MS/MS, while Abbott may reduce ONDST sensitivity, necessitating assay-specific cut-offs.

Area of Science:

  • Clinical Chemistry
  • Endocrinology
  • Laboratory Medicine

Background:

  • The 1 mg overnight dexamethasone suppression test (ONDST) is crucial for diagnosing Cushing's syndrome and evaluating adrenal incidentalomas.
  • Significant variations exist in serum cortisol immunoassay performance, yet their impact on ONDST results is under-researched.

Purpose of the Study:

  • To evaluate and compare the performance of three common immunoassay platforms (Roche Elecsys II, Abbott Alinity, Siemens Centaur) for serum cortisol measurement during ONDST.
  • To assess these platforms against a gold-standard liquid chromatography tandem mass spectrometry (LC-MS/MS) method.

Main Methods:

  • Seventy-seven anonymized patient samples from ONDSTs were analyzed using Roche, Abbott, and Siemens immunoassay platforms.
  • Results were statistically compared to a validated LC-MS/MS method, excluding samples with factors affecting immunoassay quality.

Main Results:

  • The Roche Elecsys II demonstrated minimal bias (-2.4 nmol/L) and strong correlation (y = -0.9 + 0.97x) with LC-MS/MS.
  • The Siemens Centaur also showed good agreement (bias 2.3 nmol/L, y = 1.4 + 1.07x), with minor sex-based differences.
  • The Abbott Alinity exhibited significant bias (-18.8 nmol/L) and weaker correlation (y = -11.3 + 0.88x), with notable differences between sexes.

Conclusions:

  • Method-dependent variations in serum cortisol assays significantly impact ONDST results.
  • Roche and Siemens platforms offer reliable performance comparable to LC-MS/MS for ONDST.
  • The Abbott platform may compromise ONDST sensitivity, highlighting the need for assay-specific diagnostic cut-offs.
Abstract