FRET causing misleading signal from fluorescein excited by the violet laser in flow cytometry

Louis Waeckel1, Hana Khenine2, Anne-Emmanuelle Berger1

  • 1Immunology Lab, University Hospital, Saint-Etienne, France.

Insights

A new fluorescence resonance energy transfer (FRET) interference was identified between Pacific Blue (PB) and FITC fluorochromes, excited by a violet laser. This FRET effect can cause misleading signals in multicolor flow cytometry, especially when labeling closely related markers.

Area of Science:

  • Immunology
  • Biotechnology
  • Analytical Chemistry

Background:

  • Multiple fluorochromes in immunolabeling can cause spectral interferences.
  • Fluorescence Resonance Energy Transfer (FRET) is a known phenomenon that can lead to unexpected signals.
  • Previous FRET observations in T cell clonality analysis involved PE-Cy7 and APC-Alexa750.

Purpose of the Study:

  • To report a novel FRET effect between Pacific Blue (PB) and FITC fluorochromes.
  • To investigate the cause of an unexpected signal on the V550 (Krome Orange; KrO) detector in T cell analysis.
  • To highlight the potential for FRET to cause misleading results in multicolor flow cytometry panel design.

Main Methods:

  • Fresh whole blood was stained with a panel including anti-CD3-PB, CD8-KrO, Vbeta FITC, Vbeta PE, and CD4 AA750.
  • Cells were fixed, treated for erythrolysis, washed, and analyzed on a DxFlex cytometer.
  • Data analysis was performed using Kaluza software, including spectral analysis of fluorochromes.

Main Results:

  • An unexpected CD8dim-KrO (V550) signal was observed on all Vβ FITC positive T cells when excited by a violet laser.
  • This FRET effect between PB and FITC was confirmed by its dependence on the presence of both anti-CD3-PB and anti-Vβ-FITC antibodies.
  • The FRET was also observed with anti-TCR gamma delta-FITC but not with other FITC-labeled antibodies not associated with the CD3-TCR complex.

Conclusions:

  • A FRET effect between Pacific Blue (PB) and FITC, excited by the violet laser, has been described for the first time.
  • This PB-FITC FRET can lead to misleading signals, particularly when labeling closely related markers like CD3 and TCR Vβ.
  • Careful consideration of FRET is crucial during multicolor flow cytometry panel design to avoid erroneous interpretations.