Method Validation and Establishment of Reference Intervals for an Insulin-like Growth Factor-1 Chemiluminescent

Arne Güssow1, Sabine Thalmeier1, Ruth Gostelow2

  • 1Clinic for Small Animals (Internal Medicine, Clinical Pathology and Clinical Pathophysiology), Justus-Liebig-University, 35392 Giessen, Germany.

Veterinary Sciences
|September 27, 2023
PubMed

Insights

A new chemiluminescent immunoassay (CLIA) accurately measures feline insulin-like growth factor-1 (IGF-1), offering a safer alternative to RIA for diagnosing hypersomatotropism (HS) in diabetic cats.

Area of Science:

  • Veterinary diagnostics
  • Endocrinology
  • Assay validation

Background:

  • Radioimmunoassay (RIA) for insulin-like growth factor-1 (IGF-1) is the sole method for diagnosing feline hypersomatotropism (HS).
  • Limitations of IGF-1 RIA include radiation hazards, limited availability, and high costs, necessitating alternative diagnostic methods.
  • Automated analyzers require validated assays, such as chemiluminescent immunoassay (CLIA), for routine clinical use.

Purpose of the Study:

  • To validate a CLIA for feline IGF-1 measurement.
  • To compare CLIA performance against the established IGF-1 RIA method.
  • To establish a reference interval (RI) and diagnostic cut-off for HS in diabetic cats.

Main Methods:

  • Assay validation included precision, linearity, and recovery studies for the CLIA.
  • A reference interval (RI) was determined using serum from 50 healthy adult cats.
  • Method comparison involved 68 diabetic cat serum samples, analyzed by both CLIA and RIA, with ROC analysis for cut-off determination.

Main Results:

  • The CLIA demonstrated excellent precision (intra-assay CV ≤4.7%, inter-assay CV ≤5.6%) and linearity (R² > 0.99).
  • A very high correlation (rs = 0.97) was observed between CLIA and RIA, with a mean bias of -24.5% for CLIA.
  • The established RI upper limit was 670 ng/mL, and the optimal HS diagnostic cut-off was 746 ng/mL (Sensitivity 84.4%, Specificity 97.2%).

Conclusions:

  • The validated CLIA exhibits good performance characteristics for feline IGF-1 measurement.
  • The established RI and HS diagnostic cut-off enable the use of CLIA in routine veterinary diagnostics.
  • CLIA provides a viable, non-radioactive alternative for diagnosing hypersomatotropism in diabetic cats.